Duplication prohibited
MIR 9000H
Environnement
S.A
APRIL 2011
4–9
MAINTENANCE SHEET
Serial n°MONITOR :
OPERATION SHEET 4.3.3
Subject : Checking of flow rate and ejector efficiency
FOLIO : 1/1 Frequency : 1 month
See Figure 4–1
Date
– Flow rates checking :
Verify flow rates in measurement mode, zero and span with respect to the values
written down on the control sheet, at fluid checks paragraph.
The flow rate is verified in the « Tests / Optical bench » menu.
If the flow rates are not sufficient, check status of the safety filter of the gas inlet (5).
Check also the ejector.
If the measurement flow rate is sufficient, also check the thin filter in the HOFI box or
the LCPD and the filter (optional) at the end of the sampling probe.
– Tool required
Vacuometer