3.3.2.3 MEASUREMENT
Synoptic
It presents in diagram format the internal fluid and electric data of the analyzer.
Specific “test points” inform in real time about the internal data to be analyzed. The lower part of the
screen is reserved for the display of 4 measurements.
(1) I Peltier
The following parameters are found in this menu:
Flowrate
IR source voltage
Motor
speed
Clamp
voltage
PELTIER
current
Optical bench temperature
Pressure
These values have to be compared with data shown on the control document.
An indicator appears at the top left showing the cycle under way.
M = Measurement
Z = Zero
S = Span
R = Zero Reference
A = Alarm
C = Control
Keys
Identical to the display of the “
Measurement / Instantaneous”
menu.
Apart from the 4 measurements, the « >> » key appears at the bottom right and enables displaying
the following measurements.