Duplication prohibited
MIR 9000H
Environnement
S.A
3–25
APRIL 2011
3.3.3.4 SPAN
Flow / Pressure NDIR
This screen is only activated if the NDIR module is programmed. It enables to define and linearize the
flow rate and pressure data.
CAUTION:
Modification of the parameters contained in this screen could cause
breakdowns in the analyzer or wrong measurements.
This screen enables calibrating the internal gas flow and sample pressure sensors making a
linearization of the corresponding analog inputs.
The Flow rate information shown on the
” Measurement” screens
in a graph of several stars is
defined by the two parameters:
Low range
High range
NOTE :
For a sample flow of 20 l/h and 90 l/h in zero the flowrate graph is defined between 0 and
100 l/h.
In addition, the thresholds of ”CONTROL” and “ALARM“ are programmed.
For the flow rate:
“Flow rate check” for Flow rate < Control Parameter
“Analyzer flow rate” alarm for Flow rate < Alarm Parameter
For the pressure:
“Pressure check” for Pressure < Control Parameter
“Analyzer pressure” alarm for Pressure > Alarm Parameter
The flow rate and pressure measurements are displayed in the «
Tests / Optical Bench»
menu.
STOP
Summary of Contents for MIR 9000H
Page 8: ...EnvironnementS A MIR 9000H Duplication prohibited 0 8 APRIL 2011 Page intentionally left blank...
Page 29: ...Duplication prohibited MIR 9000H EnvironnementS A 2 5 APRIL 2011 Page intentionally left blank...
Page 94: ...EnvironnementS A MIR 9000H Duplication prohibited APRIL 2011 4 2 Page intentionally left blank...
Page 97: ...Duplication prohibited MIR 9000H EnvironnementS A APRIL 2011 4 5 Page intentionally left blank...
Page 127: ...Duplication prohibited MIR 9000H EnvironnementS A APRIL 2011 5 11 INCORRECT CORRECT...
Page 142: ...EnvironnementS A MIR 9000H Duplication prohibited APRIL 2011 6 2 Page intentionally left blank...