MAINTENANCE SHEET
Monitor serial No.::
OPERATION SHEET: 4.3.10
Scope : Contamination test
PAGE : 1/1
Periodicity : 1 month
Dates
Procedure :
1/ Go to the
« TESTS
Mother board Inputs / Outputs »
.
2/ Go to
Plate
then press down the [Test] key then the [ s ] key under
Mov. Down
.
The plate moves down. The motor stops after five seconds.
3/ Press down the [
] key then go to
Filter
then press down the [Test] key then the [ s ] key
The filter ribbon moves on. The motor stops after seven seconds.
4/ Press down the [
] key, then go to
Plate,
then press down the [Test] key, then the [ s ] key under
Moving up
.
The plate moves up. The motor stops after five seconds.
5/ Press down the [
] key, then go to
P. Source
, then press down the [Test] key, then the [ s ] key
under
Sampling
.
The source is placed into sampling position. The motor stops after five seconds.
6/ Go to the
HV Geiger
line. Then, on the
ON/OFF
field, modify the parameter into
ON
and set in
action the high voltage of the Geiger counter supply
The counter is supplied with 600 V
100 V. The counting value is displayed on the below line.
7/ Check that the counting rate does not exceed 30 counts/s.
Summary of Contents for MP101M
Page 8: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 0 8 Page intentionally blank...
Page 114: ...EnvironnementS A MP101M Duplication prohibited SEPTEMBER 2013 3 58...
Page 131: ...Duplication prohibited MP101M EnvironnementS A 4 9 SEPTEMBER 2013...
Page 145: ...Duplication prohibited MP101M EnvironnementS A 4 23 SEPTEMBER 2013...
Page 163: ......
Page 184: ...EnvironnementS A ESTEL Board Duplication prohibited JUNE 2009 18 Page intentionally left blank...