Press down the [Start] key to start the calibration.
This screen enables the user to monitor the phases sequence with the « > » cursor : the device
enters into « Initialization » phase, the filter paper is fed for one step, and the pressure plate is placed
in the up position, against the filter paper. Then, the pump starts: the device passes into
« Calibration » phase, the « Reference » flow rate, preset, is displayed, as well as the measured «
Flow rate ».
If the difference between both these values is higher than 1,5 l/min, the calibration is
wrong, the instrument must be re-calibrated.
–
To do that, connect a reference flow meter to the flexible channel, and press down the
[Details] key: the
« Calibration details »
screen with all the calculation parameters is displayed.
NOTE :
The flexible channel is described in Chapter 1, section
Gas transport system, pa
ge 1-6 ,
and shown in (4) of figure 1-4.