AlcoQuant® 6020 – Breathalyser – Operating Instructions
Error display
Error display
Error display
Error display
Possible cause
Possible cause
Possible cause
Possible cause
Remedy
Remedy
Remedy
Remedy
Device errors which endanger the measuring process (red LED)
Device errors which endanger the measuring process (red LED)
Device errors which endanger the measuring process (red LED)
Device errors which endanger the measuring process (red LED)
Temperature
Temperature
Temperature
Temperature
The temperature established in the
device is beyond the specified
range
of between -5 and +50°C.
Measurement may be faulty.
Beyond the specified temperature range
the measured values established may be
subject to greater error tolerance than
specified.
The device must be switched off and
brought to within the specified
temperature range (by cooling or
heating).
Then the device must be switched on
again.
If the error occurs although the device is
evidently within the specified temperature
range, please inform Service.
Sampling system
Sampling system
Sampling system
Sampling system
An error was detected in the
sampling system
The measured value may be faulty.
Measure again.
If the error recurs, please inform Service.
Memory
Memory
Memory
Memory
An error has occurred during
reading or writing access to the
internal memory.
Measuring is not possible.
The measured value may not be displayed
correctly.
Data cannot be read correctly off the
memory and the established measured
values cannot be logged.
Measure again.
If the error recurs, please inform Service.
Alcohol sensor
Alcohol sensor
Alcohol sensor
Alcohol sensor
An error was detected on the
alcohol sensor
Measuring is not possible.
Switch the device off and then on again.
If the error recurs, please inform Service.