LCD Screens after Configuration
Start Delay
LCD displays
“
Stby
” after configuration.
Temperature Start
LCD displays
“ ”(Over), ” ” (Under) or “ ” (Between) above “
Stby
”
indicating temperature start is enabled after configuration. Then you need to
press
Start/Stop
button to active logger to measure current temperature
condition.
Auto Start
LCD displays “
Stby
”
and a
Clock
after configuration
indicating logger is
waiting for
Auto Start
on pre-set time.
Repeat Start
Repeat Start is an independent option of Startup, if enabled, you could restart Alpha data logger for
next round of monitoring with the same recording parameters of previous recording by pressing
Start/Stop button, no need to program the logger again.
“rPt”
flashes for about 5 seconds indicating you are about to re-start
recording, you are allowed to cancel restart operation at any time during
flashing.
▪
Repeat Start is only effective after recording is stopped.
Repeat Start is only available for multi-use model.
Disable Stop Button
This option is used to avoid mistaken Stop Button action, once enabled, device could not be stopped
by pressing
Start/Stop
button, recording will stop when logging cycle ends up, however, you are still
allowed to stop recording with TempCentre.