7.1.4 Flexure compensation templates that can be used in OBs
Two new templates are available in order to provide the possibility for the user to do
additional flexure compensations in case of a long OB (longer than 1h-1h15mn). This new
kind of template can be inserted between 2 science templates for instance.
In all cases, the flexures compensation is always performed at the beginning of an OB
through the acquisition template. Thus, for usual OB (shorter than 1h-1h15mn) there are no
needs to add this kind of template.
In case of slit observation, you could use the XSHOOTER_slt_AFC template. In case of IFU
observation, you could use the XSHOOTER_ifu_AFC template.
7.1.5 Science templates
Slit observations
The SEQ.AGSNAPSHOT is not available because a snapshot of the A&G camera is saved
at the end of the acquisition.
Table 22: Parameters for stare mode observations with the template
XSHOOTER_slt_obs_Stare.
XSHOOTER_slt_obs_Stare
Keyword
Range
Default Value
Label in P2PP
Free parameters
INS.OPTI3.NAME
1.0x11
UVB slit
INS.OPTI4.NAME
0.9x11
VIS slit
INS.OPTI5.NAME
0.9x11
NIR slit
DET1.WIN1.UIT1
0..36000
UVB Exposure Time (s)
DET1.READ.CLKDESCR
100k/1pt/hg
UVB readout mode
DET2.WIN1.UIT1
0..36000
VIS Exposure Time (s)
DET2.READ.CLKDESCR
100k/1pt/hg
VIS readout mode
DET3.DIT
0..36000
NIR Detector Integration
Time (s)
DET3.NDIT
1..20
1
number of DITs
SEQ.NEXPO.UVB
0..100
1
UVB number of exposures
SEQ.NEXPO.VIS
0..100
1
VIS number of exposures
SEQ.NEXPO.NIR
0..100
1
NIR number of exposures
Fixed Values
INS.MODE
SLITSPEC,
IFUSPEC
SLITSPEC
Instrument Mode
SEQ.AGSNAPSHOT
T, F
F
Take an acquisition image
before science exposures