Doc:
Issue
Date
Page
VLT-MAN-ESO-14650-4942
P96
24.06.2015
56 of 161
ESO, Karl-Schwarzschild-Str. 2, 85748 Garching bei München, Germany
About the UVB CCD, the recent tests show that the detector does not have remnants after
arc exposures of 6 to 300s with the 1x1 binning and the normal readout mode, 3s exposure
in the 1x2 binning for the normal readout mode, and 1s exposure for the 2x2 binning in the
normal readout mode.
However it was observed some remnants after the observation during 300s of a very bright
star that saturated the UVB detector.