F1.2 Calibration
F1.2.1 Calibration
1)
Start with Enter-Key
2)
Display
E
-S
cl
,
empty
scale
and
confirm with Enter-Key
3)
Display counts down 10 … 0 to
capture Zero
4)
Display
will
s
ay
Add Load
5)
Confirm with Enter-Key
6)
With Tare-Key or Zero-Key the number can be changed
like described under number input.
7)
Put load on scale and confirm with Enter-Key
8)
Display will count down 10..0. If scale is not stable after
30 seconds it will time out and display an error code.
With enter-key go back to calibration
beginning.
9)
If calibration was successful it will display done for 2
seconds.
10)
Confirm with Enter-Key.
Supervisor mode: Go to function block F7
Administrator mode: Go to function block F2
F1.1.2 Capacity
Selection:
3…4,000(2,000 is default setting)
F1.1.3 Increment
F1.1 Capacity & Increment
F1.1.1 Units
Selection:
- kg(default setting)
- lb
11.3. Function block Setup
11.3.1. F1 - Calibration
11.3.2. F2 Function block Scale
F2.1 Approval
Selection:
- No, not approved (default setting)
- OIML
F2.2 GEO
Selection:
0…31(16 is default setting)
F2.3 Zero
F2.3.1 Auto Zero Maintenance
Selection:
off, 0.5d, 1d, 3d (0.5d is default setting)
F2.3.2 Power up Zero (based on Cal Zero)
Selection:
off, 2%, 10%, 20 % (10% is default setting)
F2.3.3 Pushbutton Zero (based on Cal Zero)
Selection:
off, 2%, 10%, 20% (10% is default setting)
F2.4 Tare
F2.4.1 Auto Tare
- NT P
E
Selection:
On/Off (Off is default setting)
F2.4.2 Auto Clear Tare
Selection:
On/Off (Off is default setting)
F2.4.3 Tare Interlock
Selection:
On/Off (Off is default setting)
F2.4.4 Auto Tare threshold(only active if F2.4.1 is On)
Selection
0…FS (0 is default setting)
F2.4.5 Auto tare reset threshold(only active if F2.4.1 is On)
Selection
0…FS (0 is default setting)