EasyMax Voice III 

Self-Monitoring Blood Glucose System 

 

 

Quick Reference Guide 

 

 

 

 
 

 

Summary of Contents for EasyMax Voice III

Page 1: ...1 EasyMax Voice III Self Monitoring Blood Glucose System Quick Reference Guide...

Page 2: ...pen the battery door on the back of the meter by pushing the tab in the direction of the arrow and pulling the door up 2 Insert two batteries You will hear opening music 3 Put the battery door back in...

Page 3: ...on to select ES En English ES Espa ol 2 And press power button to confirm the setting 3 Repeat step 1 and 2 to set the volume 1 The priority language of the blood glucose meter is set in English 2 The...

Page 4: ...te in your meter is important if you intend to use the meter memory 1 Press M button and C button to set the year Press power button to confirm the setting 2 Repeat step 1 to set the date 3 Repeat ste...

Page 5: ...a flat surface like a table 4 Remove the control solution bottle cap and wipe the tip of the bottle with a tissue 5 Squeeze the bottle until a tiny drop forms at the tip of the bottle 2 The meter turn...

Page 6: ...within the range printed on the test strip bottle 9 Press power button to repeat the test result Or you can also press C button or M button to set the volume before repeating the result 10 Remove the...

Page 7: ...The Lancing Device You must first load the lancet into the lancing device to get it ready for use 1 Unscrew the Cap 2 Insert the lancet into the lancing device firmly then twist off the protective co...

Page 8: ...average and 4 5 for thick or calloused skin 6 Set the lancing device aside until later in the test 5 Pull on the sliding barrel of the lancing device until it clicks and then release Now the lancing d...

Page 9: ...ing 3 When the blood drop flashes on the display the meter will tell the instruction Running A Blood Glucose Test With Blood From Your Fingertip 1 Wash your hands with soap and warm water Rinse and dr...

Page 10: ...indow Otherwise an inaccurate reading may result 7 The meter starts to count down from 5 seconds and speaks the result 8 Press power button to repeat the test result Or you can also press C button or...

Page 11: ...ble tip against the forearm 3 Press the trigger button to activate the lancing device Hold the device against forearm and increase pressure until the blood sample size is sufficient Running A Blood Gl...

Page 12: ...he cap 3 Pointing the lancing device toward a container for sharp or biohazard material slide the ejection button down to release the covered lancet into the container 2 Without touching the used lanc...

Page 13: ...individual test result 4 To delete a test result press C button for more than 3 seconds and display shows dEL Then press power button to confirm the deletion 5 Press C button to keep reviewing the re...

Page 14: ...n error message If you have any concerns please contact your local dealer Humidified Used strips Action Replace with a new strip Low power Action Replace with new batteries System fault Action Replace...

Page 15: ...still the same try a control solution test and if the control solution test falls within the correct range please contact your healthcare professional The test result is lower than 1 1 mmol L 20 mg d...

Page 16: ...test continues Action Relocate the meter to a location with temperature between 10 C 40 C 50 F 104 F The Lt and thermometer icon appears Temperature is too low outside the required range of 10 C 40 C...

Page 17: ...mail info epsbio com tw Website http www epsbio com tw 0537 Reference American Diabetes Association Standards of Medical Care for Patients with Diabetes Mellitus Diabetes Care 25 2002 p S37 Obelis s...