Page 58
EPS Stromversorgung GmbH
Alter Postweg 101 • 86159 Augsburg
Germany
PSI 9000 3U Series
Fon: +49 821 / 570451-0
Fax: +49 821 / 570451-25
3.9
The function generator
3.9.1 Introduction
The built-in function generator is able to create various signal forms and apply these to the set value of voltage
or current.
In manual operation, all generator functions are available for access on the front panel. In remote control, only the
customisable arbitrary generator and a XY function are available. The arbitrary generator can replicate all manually
serviceable functions, except UI and IU. For those, the XY function is assigned.
The following functions are retrievable, configurable and controllable:
Function
Usable
on
Short description
Sine wave
U, I
Sine wave generation with adjustable amplitude, offset and frequency
Triangle
U, I
Triangular wave signal generation with adjustable amplitude, offset, gain and decay times
Rectangular
U, I
Rectangular wave signal generation with adjustable amplitude, offset and duty cycle
Trapezoid
U, I
Trapezoidal wave signal generation with adjustable amplitude, offset, rise time, pulse
time, fall time, idle time
DIN 40839
-
Simulated automobile engine start curve according to DIN 40839 / EN ISO 7637, split
into 5 curve sequences, each with a start voltage, final voltage and time
Arbitrary
U, I
Generation of a process with up to 100 freely configurable steps, each with a start and
end value (AC/DC), start and end frequency, phase angle and total duration
Ramp
U, I
Generation of a linear rise or fall ramp with start and end values and time before and
after the ramp
UI-IU
-
Table (.csv) with values for U or I, uploaded from a USB flash drive, as well as func
-
tions to simulate a solar panel (PV function) or fuel cells (FC function)
Whilst R mode is activated, access to the function generator is not available.
3.9.2 General
3.9.2.1 Limitations
The function generator is not accessible, neither for manual acces, nor for remote control, if
•
master-slave mode has been activated and the device was configured as slave.
•
resistance modes (R/I adjustment mode, also called UIR mode) is active.
3.9.2.2 Resolution of X (time) and Y (amplitude)
The device can set 4096 steps between 0 ... 100% of nominal value. The intervals to create a linear or other rise/
fall are calculated depending on the amplitude and then set.
If the amplitude is very low and the time long, only few intervals will be set as otherwise many identical values will
be set one after another, generating a staircase effect.
3.9.2.3 Possible technical complications
Operation of switching mode power supplies as a voltage source can, when applying a function to the output volt
-
age, lead to damage of the output capacitors due to continuous charging/discharging which causes overheating.
Furthermore the actual voltage progression may differ from what’s expected.
3.9.3 Method of operation
In order to understand how the function generator works and how the value settings interact, the following should
be noted:
The device operates, including in function generator mode,always with the three set values U,I and P.
The selected function can be used on
one
of both the values U or I, the other two are then constants and have
a limiting effect. That means if, for example, a voltage of 10 V is set for the DC output, a load is connected and
a sine wave function should operate on the current with an amplitude of 20 A and offset 20 A, then the function
generator will create a sine wave progression of current between 0 A (min) and 40 A (max), which will result in an
output power between 0 W (min) and 400 W (max).The output power, however, is limited to its set value. If this
were 300 W then, in this case, the current would be limited to 30 A and, if clamped to an oscilloscope, it would be
seen to be capped at 30 A and never achieve the target of 40 A.