A3 Size Color Flat Bed Scanner GT-30000
Revision A
ADJUSTMENTS
Adjustments Using the Special Software
60
4. The current setting data rear from the EEPROM will be displayed in the
respective areas.
5. The EEPROM data can then be edited.
5.3.3.2 Japanese/English Display Select Button
Click this button to toggle the display language for the dialog box between
Japanese and English.
5.3.3.3 Changing the Scanner Model
1. Select the scanner model from the scanner model selection list.
2. Click
Write
.
5.3.3.4 Adding a new scanner model
1. If you would like to add a new scanner model, enter the name of the new
scanner model in the additional model display area.
2. After entering the model name, click
Add
. The new scanner model will then
be added to the end of the scanner model selection list.
The new scanner model can then be selected in the same way as other
scanner models.
5.3.3.5 Deleting the Selected Scanner Model
1. Select the scanner model that you would like to delete so that the model
name appears in the additional model display area.
2. After checking that the selected scanner model is correct, click
Delete
.
3. The selected scanner model will then be deleted from the scanner model
selection list.
5.3.3.6 Resetting the number of scans to zero
1. Click 0 Clear. The 0 Clear confirmation message box will be displayed.
2. To reset the value, click
OK
. To cancel the operation, click
Cancel
. The
data will only be reset to zero if you click
OK
.
C A U T I O N
!
Due to the identity of the program, if you input some
data to any area and click "write" button, the life
counter of scanning frequency and ADF paper
sheets fed will be reset automatically. Therefore, it is
recommended to take a memo before you edit any
area.
!
Prepare an note to write down each scanners life
counter when you service or repair a scanner.
C H E C K
P O I N T
Be sure to write the data to the EEPROM after each
editing change has been made.
C H E C K
P O I N T
Be sure to set the scanner model after replacing the
main circuit board.
C A U T I O N
Note that when the number of scans (cumulative
life) is reset to zero, the previous value will not be
retained.
Summary of Contents for GT-30000 Series
Page 9: ...C H A P T E R 1 PRODUCTOUTLINE ...
Page 22: ...C H A P T E R 2 OPERATINGPRINCIPLES ...
Page 29: ...C H A P T E R 3 TROUBLESHOOTING ...
Page 37: ...C H A P T E R 4 DISASSEMBLY ASSEMBLY ...
Page 55: ...C H A P T E R 5 ADJUSTMENTS ...
Page 65: ...C H A P T E R 6 MAINTENANCE ...
Page 67: ...C H A P T E R 7 APPENDIX ...
Page 82: ...Model ES 9000H GT 30000 Board PBA MAIN Rev 1 Sheet 1 3 ...
Page 83: ...Model ES 9000H GT 30000 Board PBA MAIN Rev 1 Sheet 2 3 ...
Page 84: ...Model ES 9000H GT 30000 Board PBA MAIN Rev 1 Sheet 3 3 ...
Page 85: ...Model ES 9000H GT 30000 Board PWS POWER Rev A Sheet 1 1 ...
Page 86: ...Model ES 9000H GT 30000 Board PBA Connect Rev 1 Sheet 1 1 ...