ELite 3, User Manual
Page 21/72
2.3.2.
Turning the Instrument OFF
DO NEVER turn off the analyzer by simply flipping the power switch on the rear
panel. Doing so may result in erroneous operation during later use. It can be so,
because the instrument uses diluent. This liquid is an isotonic saline solution
containing salt. Would it not be washed out of special units of the instrument or would
chambers not be filled with this solution may lead to dust condensation or salt build
up.
Therefore always follow the instructions below when switching the instrument off.
In Main menu, select SHUTDOWN. The following screen appears.
EXIT
SHUT DOWN
Select Shutdown.
The analyzer will perform the necessary
steps to prevent failure to the pneumatic
system, and then gives a tone indicating
that it is safe to shut it off.
Turn off power using the rear power switch.
Turn off the instrument using the power switch on the rear panel. The ‘OFF’ position
is marked by ‘O’ symbol.
2.3.3.
Preparing for shipment
Use the second item in the Shutdown menu when the instrument is to be shipped or
left unused for a longer time (more than 1 week). The instrument will ask you to use
the cleaning tube kit and 100ml of distilled water.
Follow the instructions appearing on the display.
SHUTDOWN
Preparing for shipment (2)
Remove tubing connectors, so the system
can drain itself.
Leave the waste connector attached.