ELite 5, User Manual
Page 116 of 159
Chamber
-20 - 20mBar
Noise/Pulse
pls/5sec
0 - 2000 pulses
20000pls
19990 - 20050 pulses
HGB head
HGB Dark
0 - 3000 pulses
HGB light
3000 - 60000 pulses
TCU temperature
Actual
Ref- 0.2 °C
Sink
0 - 55°C
Laser parameters
Ref-10%
Laser power 32.0 – 128.0
Laser DC level
Laser off
0 - 0.05mV
Laser on
0.10 - 0.45mV
HGB head
HGB Dark
0 - 3000 pulses
HGB light
3000 - 60000 pulses
Electrode
Voltage
45 - 55V
Current
620 – 680
μ
A
Offset
-3.0 - 1mV
Reagent lock
Status: UNLOCKED / LOCKED / NA
Available measurements: none
Battery /Power supply
Battery voltage
2.7 - 3.3V
+12V
11.4 - 12.6V
-12V
-12.6 - -11.4V
Temperature
WBC pre-heater:
none (LSDACQ v3.0),
34.0 -38.0 (LSDACQ v3.1)
LSDACQ: none
If any of the values falls outside the above limits, the SW is going to list the errors on the self
test report screen.