ELite 5, User Manual
Page 133 of 159
Cleaning:
Perform cleaning / hard cleaning if:
High blank results experienced;
Clogging related flags are added systematically to measurement results;
After service actions, service level maintenance;
Measurement results, QC results are drifting (perform hard cleaning);
Clog formation in the tubing (first remove the plugging, contamination);
Reagent connected to wrong place, wrong reagents were used.