7.4.3 Automatic Mode
Only human samples in closed sample tubes can be processed in automatic mode from the
Autosampler. QC samples and open tubes must be processed in manual mode.
After starting up the ‘ELite 5’, select the ‘Measure’ quick link. A new blank must be processed
if this is the first time a new measurement panel is opened since starting up, or if an earlier
blank has expired due to inactivity. An accepted blank measurement is required before
starting automatic sample processing.
Open the AS panel by using the “AS” quick link at the top of the screen.
From the AS panel, you can initiate automated sample measurement in 3 modes:
•
Full Scan mode:
-
The Autosampler will scan the whole sample tray
-
The same sample mode will be used for all the samples
-
The sample ID is generated from the sample tube barcode (if available)
-
It is not possible to assign a patient to automatically processed samples. The
default patient will be used for all samples.
•
Free List mode:
-
The samples are defined in a list.
-
The following parameters can be individually defined per sample:
Sample ID
Patient
Sample mode
-
The samples should be placed in the same order in the sample tray as the
samples defined in the list.
Figure 41. Autosampler Panel
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...