ELite 5, User Manual
Page 64 of 159
Sample (in position 4) has bar code and cap. Sample measurement is finished.
Sample (in position 1) has no barcode but has cap. Sample measurement is
finished.
Sample (in position 2) has no cap. Sample is ignored.
7.4.3.2 Full Scan Mode
Full Scan mode is suitable for laboratories where:
•
There are batches of samples with the same sample mode,
•
Barcode labels are in use, and
•
An LIS system rather than the ‘ELite 5’ is used to associate patients with sample
results.
In Full Scan mode it is possible to add more samples to the batch after starting the
automated measurement. As usual, an accepted blank is necessary before starting
automatic sample processing.
Figure42. Full Scan Automatic Processing Mode
On the Full Scan mode panel you can:
•
Return to the AS panel to select a different type of automated measurement by using
the ‘Back’ function.
•
Select a common measure mode for all samples in the tray.
•
Swap between the tray and list views by using the ‘View list’ and ‘View tray’ functions.
•
Start the automated measurement by clicking or tapping on the ‘Start’ button on the
GUI.