The analyzer displays a blank result screen and begins sample measurement of the
calibration material. Wait for the screen to be populated with results when sample
measurement is complete.
•
Click or tap the ‘Accept’ button if the measurement run is acceptable.
•
Click or tap the ‘Start next’ button to begin the next calibration run. Selecting ‘Start
next’ without accepting a calibration run discards the run. Selecting ‘Abort’ will abort
the entire calibration procedure and discards all calibration results.
•
The calibration procedure ends when three or seven accepted calibration runs are
completed. The calibration factors are calculated and presented in the calibrator
factor. The previous calibration factors and the coefficient of variation (CV) of the
sample runs for each parameter are also displayed.
•
Select ‘Accept’ to replace the current calibration factors with the accepted factors.
Selecting ‘Discard’ eliminates the entire calibration process.
Figure 68. Calibration Factor Panel
The automatic calibration process can be skipped by selecting the ‘Manual’ calibration mode
on the calibration panel and clicking or tapping the ‘Next’ button proceeds directly to the
calibration factor panel for manual entry of factor values. Manually entered factor values can
be accepted in the same way as calculated factors from calibration sample runs, and are
treated by the ‘ELite 5’ in the same way.
10.1.1 View Calibrations
The view calibration features is helpful for reviewing the calibration history for the analyzer.
Increasing or decreasing calibration value trends over time may indicate a problem that
requires service intervention.
Summary of Contents for ELite 5
Page 13: ...ELite 5 User Manual Page 13 of 159 Figure 1 ELite 5 with Autosampler Space Requirements...
Page 27: ...ELite 5 User Manual Page 27 of 159 Figure 9 4DIFF Scatter Diagram...
Page 142: ...ELite 5 User Manual Page 142 of 159 20 3 Printed Report Formats...
Page 143: ...ELite 5 User Manual Page 143 of 159...
Page 153: ...ELite 5 User Manual Page 153 of 159 20 5 Fluidic System...