PWC 170 = heart rate upper limit of 170, trained test person
On PWC 130, the initial strain is 25 W; on PWC 150 and PWC 170, the initial strain is
50 W each. The strain increases by 25 W (PWC 130) and by 50 W (PWC 150, PWC
170) every two minutes. An assessment will only be carried out after the set heart rate
has been reached and the set strain step has been ended.
The test result will be a Watts/kg value. With this value and the PWC classification
table (see Chapter A8 "Test evaluation"), you can assess your actual fitness level.
How to carry out a test:
1.
Push the PLUS/MINUS buttons until the programme TEST is selected. Confirm
your choice with the START button.
2.
You are now within the "TEST" submenu. Choose the desired test programme
with the PLUS and MINUS buttons and confirm this with START.
3.
Now choose your gender with the PLUS and MINUS buttons and again confirm
your choice with START (M= male; F= Female).
4.
Press the PLUS and MINUS buttons to enter your age and press the START
button.
5.
Now enter your body weight. For this purpose, press the PLUS and MINUS
buttons (default weight: 70 kg). Confirm this with START.
6.
The display now shows the RESTING PULSE ASSESSMENT mode. Your
current resting heart rate will be assessed. This takes about 15-20 seconds.
Your resting heart rate is displayed.
7.
If the assessment is successful, the test will start after a few seconds. The
display changes to operating mode and starts with the first strain step.
8.
As soon as your heart rate exceeds the set upper limit (130, 150, 170), the test
ends automatically after completion of the strain level.
9.
The training parameters remain on the display after the test is completed. By
pushing the STOP button one more time, you will return to the main menu.
7.1.8
ECG Choice
An external connection (RS 232, serial interface) is standard equipment on the CYCLE
457 MED. The interface cable is not included in the delivery.
Danger! Some ECG devices cannot be connected with the CYCLE 457 MED. Call
0049-633-24610 for more information.
Control of the ergometer by external devices:
Use the interface cable to connect the bike ergometer with the external device (ECG,
Technical and optical modifications as well as misprints reserved -
© 2011
by ERGO-FIT GmbH & Co. KG
4 1