Electromagnetic Interferrence Immunity CARDIO LINE 4000 SP, 4000 S SP
ERGO-FIT products are intended for use in the following environments. Please make sure to
only use the product in appropriate environments.
Interference immunity
test
IEC 60601-
test level
Conformity Electromagnetic
environment
Discharge of static
electricity in comliance
with IEC 61000-4-2
+/- 6 kV
contact
discharge
+/- 8 kV
air discharge
6 kV
8 kV
Wooden or concrete floor
or floor to be laid out with
ceramic tiles. If the floor
is laid out with synthetic
materials, the relative
humidity must be higher
than 30 %.
Electrical fast transient
bursts in compliance with
IEC 61000-4-4
n/a
The supply power quality
should comply with that
of a business or hospital
environment.
Surges in compliance
with IEC 61000-4-5
n/a
The supply power quality
should comply with that
of a business or hospital
environment.
Voltage dips, brief
voltage interruptions
and voltage fluctuation
in compliance with IEC
61000-4-11
n/a
The supply power quality
should comply with that
of a business or hospital
environment.
Magnetic filed for supply
frequency (50/60 Hz)
in compliance with IEC
61000-4-8
Magnetic fields of the
supply frequency should
comply with those of
a business or hospital
environment.