24
ergoselect 1
6 Operation
6.9.8 RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and confirm with
Select
.
Using the arrow keys, change the value and save the new
value with
Select
.
Note
The limits selected in this menu only apply to the
load range between 6 and 150 watts. At higher
loads the RPM limits automatically adapt to the
respective load ranges:
Load (watts) Green RPM range (1/min)
6 – 150
54 – 64 (adjustable)
151 – 250
61 – 62
251 – 350
71 – 72
351 – 450
81 – 82
451 – 550
91 – 92
551 – 650
101 – 102
651 – 750
111 – 112
751 – 850
121 – 122
851 – 950
> 130
951 – 999
> 135
RPM
Min.
0 ...
70:
54
/min
Max.
50 ...
130:
65
/min
Select
Figure 6 – 29: Setting the RPM limit values
6.9.9 HR Belt Number
If the test subject wears a chest strap to measure the heart
rate during the training, the corresponding chest strap
number must be entered here. You will find the number on
the back of the housing. It is the unique identifier for this
particular strap.
With keys
↑
↓
, you toggle between the menu screens.
Select
HR belt no.
HR belt
00037942
Figure 6 – 30: Setting the HR belt no. – screen 1
When the HR belt option has been selected (see figure 6 – 30),
you can set the first digit of the number of up to 8 digits
by pressing key
↓
. The desired numeral is selected with
the arrow keys and the selected numeral is saved with the
Select
key. Repeat these steps for each numeral until the
complete number has been entered.
Select
HR belt no.
HR belt
00037942
Figure 6 – 31: Setting the HR belt no. – screen 2