32
ergoselect 1
11 Electromagnetic Compatibility EN 60601-1-2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ergoselect 1 ergometer is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that the ergoselect 1 ergometer is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level Electromagnetic Environment –
Guidance
Electrostatic discharge (ESD) to
EN 61000‑4‑2
± 8 kV contact
± 16 kV air
± 8 kV
± 16 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast transient / burst
to EN 61000‑4‑4
± 2 kV for power
supply lines
± 1 kV for input and
output lines
± 2 kV passed
Mains power should be that of a typical
commercial or hospital environment.
Surge to EN 61000‑4‑5
± 1 kV differential mode
± 2 kV common mode
± 1 kV
N. A.
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short interrup‑
tions and voltage variations
on power supply input lines to
EN 61000‑4‑11
to EN 61000‑4‑11
passed
Mains power should be that of a typical
commercial or hospital environment.
If the user of the ergoselect 1 ergometer
requires continued operation during
power mains interruptions, it is rec‑
ommended that the ergoselect 1 ergometer
be powered from an uninterruptible
power supply or a battery.
Power frequency (50/60 Hz)
magnetic field to EN 61000‑4‑8
30 A/m
50 Hz
passed
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hos‑
pital environment.
The ergoselect 1 ergometer has no compo‑
nents susceptible to magnetic fields.
Note:
UT is the a.c. mains voltage prior to application of the test level.