8 Control Terminal P
38
ergoselect 4 / 5
The stored test protocols available for selection will be
displayed. There are five fixed protocols (protocols 1 to 5,
(see chapter 13.5 “Exercise Test Protocols” on page 87),
whereas protocols 6 to 15 are user-configurable.
The protocol menu provides an overview of the test phases.
Example:
50 W / 2 min / 25 W
means: Basic load of 50 W
Stage time of 2 min
Load stage of 25 W
Use the softkeys on the right and left (
↑ ↓
) to position the
bar cursor on one of the protocols and confirm the selec-
tion with
Select
.
Protocols
1. WHO
2. BAL
3. Hollmann
4. STD. France
5. Standard
Select
Figure 8 – 9: Selecting an exercise test protocol
The exercise test is started with the
Start
key, a blood
pressure measurement at rest may precede the test
(depending on the selected exercise test protocol).
When the basic load appears on the display (after approx.
15 seconds or upon termination of the blood pressure
measurement) and the patient's RPM indicator blinks, the
patient should start pedaling.
0
Watt
--
min
0
/ min
---
/
---
mmHg
0
♥ /
min
Start
Figure 8 – 10: Initial exercise test screen
The internal protocol will now control the entire exercise
test – the display always indicates the current values.
With the +5 W and –5 W keys, the current load can be
changed any time (in increments of +/–1 W up to +/–25 W,
as configured).
120
Watt
15
min
76
/ min
138/96
mmHg
122
♥ /
min
Recovery
+ 5 W
– 5 W
Figure 8 – 11: Display during the exercise test