8 Control Terminal P
ergoselect 4 / 5
39
Terminating an Exercise Test
The exercise phase can be terminated manually at any time
with the
Recovery
key.
The load will immediately be reduced to 25 watts, but a
higher or lower value can be selected manually.
It is recommended that the patient continue to pedal in the
recovery phase.
The
End
key in the middle will terminate the test.
120
Watt
15
min
76
/ min
138/96
mmHg
122
♥ /
min
End
+ 5 W
– 5 W
Figure 8 – 12: Recovery phase
8.2.3 Manual
Use the softkeys on the right and left (
↑ ↓
) to position
the bar cursor on
Manual
and confirm the selection with
Select
.
In this operating mode the user controls the entire exercise
test by selecting the loads, stage times and by initiating
blood pressure measurements.
PC Mode
Ergometry
Manual
Settings
Select
Figure 8 – 13: Main menu
The exercise test is started with the
Start
key, afterwards
the load can be set and changed with the +5 W and –5 W
keys (in increments of +/–1 W up to +/–25 W, as con-
figured).
Blood pressure measurements can be initiated with
.
0
Watt
--
min
0
/ min
---
/
---
mmHg
0
♥ /
min
Start
+ 5 W
– 5 W
Figure 8 – 14: Initial screen of a manual exercise test
Terminating the Manual Mode
The exercise test can be terminated manually at any time
with the
End
key located in the middle.
The load will immediately drop to 0 watt.
There is no recovery phase in the manual mode.
120
Watt
15
min
76
/ min
138/96
mmHg
122
♥ /
min
End
+ 5 W
– 5 W
Figure 8 – 15: Display during the exercise test
Summary of Contents for ergoselect 4
Page 1: ...ergoselect 4 5 Ergometer Operator s Manual 201000433000 Version 2021 06 01 Rev 04 English...
Page 2: ......
Page 95: ......