8 Control Terminal P
ergoselect 4 / 5
45
Note
The limits selected in this menu only apply to the
load range between 6 and 150 watts. At higher
loads the RPM limits automatically adapt to the
respective load ranges:
Load (watts)
Green RPM range (1/min)
6 – 150
54 – 64 (adjustable)
151 – 250
58 – 65
251 – 350
68 – 75
351 – 450
78 – 85
451 – 550
88 – 95
551 – 650
98 – 105
651 – 750
108 – 115
751 – 850
118 – 125
851 – 950
> 125
951 – 999
> 130
Pulse Display
The pulse readout on the display can be turned off.
Display
Pulse Display
Select
On
Off
Figure 8 – 33: Setting the pulse readout
HR Belt Number
If the test subject wears a chest strap to measure the heart
rate during the training, the corresponding chest strap
number must be entered here. You will find the number on
the back of the housing. It is the unique identifier for this
particular strap.
With keys
↑
↓
, you toggle between the menu screens.
HR belt no.
Select
00066300
HR belt
Figure 8 – 34: Setting the HR belt no., screen 1
When the HR belt option has been selected (see figure 8 – 34),
you can set the first digit of the number of up to 8 digits
by pressing key
↓
. The desired numeral is selected with
the arrow keys and the selected numeral is saved with the
Select key. Repeat these steps for each numeral until the
complete number has been entered.
Select
0
0 0 6 6 3 0 0
HR belt no.
HR belt
Figure 8 – 35: Setting the HR belt no., screen 2
Summary of Contents for ergoselect 4