9 Control Terminal T
ergoselect 4 / 5
51
With the [ Edit ] key, you can modify each protocol para
-
meter. The new inputs overwrite the existing values.
Description
Protocol 6
Type
Step
Edit
Basic Load:
30 W
Stage Time:
2 min
Stage Rate:
25 W
Recovery Time:
2 min
Recovery Load:
25 W
Figure 9 – 16: Ergometry – editing protocol parameters, screen 2
In the configuration menu, the following parameters can
be edited:
• the basic load (from 6 to 100 W),
• the stage time (form 1 to 30 min),
• the stage rate (increment, from 1 to 400 W).
2.
Description
Protocol 6
Type
Step
Basic Load
30 W
Stage Time
2 min
Stage Rate
25 W
Figure 9 – 17: Ergometry – editing protocol parameters, screen 3
With the [
2. ] key, you proceed to the next menu level
where you can edit these parameters:
• the recovery load (from 6 to 100 W) and
• the recovery time (from 1 to 30 min).
With the [
1. ] key, you return to the previous screen.
1.
Recovery Load
25 W
Recovery Time
2 min
Figure 9 – 18: Exercise test protocol, basic load, screen 1
Touch a light gray field, e.g., at Basic Load: a submenu with
input field and numeric keypad opens. You can enter values
directly via the numeric keypad.
Confirm your inputs with the [ ] key. To cancel the input,
press the [ ] key.
The other parameters can be edited in the same way, they
will overwrite the current values.
2.
Basic Load
30 W
Stage Time
2 min
Stage Rate
25 W
Description
Type
Step
Protocol 6
Anfangslast
W
(6 W ... 100 W)
(6 W ... 100 W)
11
22
33
44
55
66
00
77
88
99
Description
Protocol 6
Basic Load
W
Figure 9 – 19: Exercise test protocol, basic load, screen 2