Pulse
During a heart-rate controlled training session, the load
increases gradually until the specified training heart rate
is reached. If the patient's heart rate exceeds the training
heart rate, the load will be reduced
The entered maximum training load should be determined
individually for each patient.
The entered heart rate should be determined individually
for each patient.
In Setup, further settings for regulation of the training
parameters can be configured for the "Pulse" training
(see also figure 9 – 79 on page 75).
2.
Description
Heart Rate
Type
Warmup Load
Warmup Time
Pulse
Belt
2
Pulse
20 W
2 min
Figure 9 – 26: Protocol type "Pulse" – level 1
Warning
Patient Hazard
Heart rate measuring systems may be inaccurate.
There is a risk of wrong or missing heart rate data
leading to inadequate therapy and/or overexertion.
Caution
If the heart rate signal is lost during a pulsecon
trolled training or the displayed value is 0, the load
will gradually be reduced to 0 watt.
1.
Training Time
20 min
Heart Rate
100 1/min
Training Load max.
50 W
Recovery Load
20 W
Recovery Time
3 min
Figure 9 – 27: Protocol type "Pulse" – level 2
W
200
150
100
50
0
0
5
10
15
20
25
1/min
200
150
100
50
0
76
1/min
98
%
15:00
mm:ss
65
1/min
25
W
mmHg
118/96
End
const.
Stop
– 5 W
+ 5 W
Figure 9 – 28: Example of a heart rate controlled training