13 Technical Specifications
ergoselect 4 / 5
87
13.4 ECG Module
[Option A]
Algorithm for calculation of the heart rate using the
ECG signal
The heart rate is calculated with the Pan Tompkins algorithm for
QRS detection.
Type of protection
BF
Input resistance
> 2.5 MΩ
Frequency range
0.65 to 40 Hz
Measuring range
± 10 mV
Common mode rejection
> 100 dB
Noise level
≤ 20 µV
AD Converter
12-bit
Sampling rate
250 Hz
13.5 Exercise Test Protocols
Protocol
Basic Load
[W]
Stage Time
[min]
Load Stage
[W]
Recovery Load
[W]
Recovery Time
[min]
1. WHO
25
2
25
25
99
2. BAL
50
3
50
25
99
3. Hollmann
30
3
40
25
99
4. STD France
30
3
30
25
99
5. Standard
20
1
25
25
99
6. – 15. (user-configurable)
25
2
25
25
99
Adjustment Range
20 – 100
1 – 30
1 – 400
20 – 100 (*)
1 – 99
(*) With Control Terminal P, the recovery load is fixed at 25 W.
13.6 Test Protocols (control terminal T only)
Protocol
Basic Load
[W]
Duration
[sec]
Load Increment
[W]
Stage Time
[sec]
Recovery Load
[W]
Recovery Time
[min]
Ramp test
0
120
25
10
25
99
PWC-130 (*)
25
0
25
120
25
99
PWC-150 (*)
50
0
25
120
25
99
PWC-170 (*)
50
0
50
120
25
99
(*) the protocol advances to the recovery phase as soon as the target heart rate (130/150/170) is reached
Summary of Contents for ergoselect 4
Page 1: ...ergoselect 4 5 Ergometer Operator s Manual 201000433000 Version 2021 06 01 Rev 04 English...
Page 2: ......
Page 95: ......