14 Electromagnetic Compatibility EN 60601‑1‑2
ergoselect 4 / 5
89
14 Electromagnetic Compatibility EN 60601-1-2
Changes or modifications to this system not expressly
approved by ergoline could cause EMC issues with this or
other equipment.
This system is designed to comply with applicable regula-
tions regarding EMC.
Its compliance with these requirements has been verified.
It needs to be installed and put into service according to
the EMC information stated as follows.
Warning
RF Interference
Use of portable phones or other radio frequency
(RF) emitting equipment near the system may
cause unexpected or adverse operation.
Caution
Equipment Malfunction
The equipment or system should not be used adjacent
to, or stacked with, other equipment. If adjacent or
stacked use is necessary, the equipment or system
should be tested to verify normal operation in the configuration
in which it is being used.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ergoselect ergometer is intended for use in the electromagnetic environment specified below. It is the responsibility
of the customer or user to ensure that the ergoselect ergometer is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions to EN 55011
Group 1
The ergoselect ergometer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions to EN 55011
Class B
The ergoselect ergometer is suitable for use in all estab-
lishments, including domestic and those directly con-
nected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions to EN 61000-3-2
Class A
Voltage fluctuations/flicker emissions
to EN 61000-3-3
Complies