14 Electromagnetic Compatibility EN 60601‑1‑2
90
ergoselect 4 / 5
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ergoselect ergometer is intended for use in the electromagnetic environment specified below. It is the responsibility
of the customer or user to ensure that the ergoselect ergometer is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level Electromagnetic Environment –
Guidance
Electrostatic discharge
(ESD) to EN 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV
± 15 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast transient/
burst to EN 61000-4-4
± 2 kV for power supply lines
± 1 kV for input and output lines
± 2 kV passed
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge to EN 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV
N/A
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short inter-
ruptions and voltage
variations on power
supply input lines to
EN 61000-4-11
< 5 % UT
(>95 % dip in UT) for 0.5 cycles
< 5 % UT
Mains power quality should be that
of a typical commercial or hospi-
tal environment. If the user of the
ergoselect ergometer requires con-
tinued operation during power mains
interruptions, it is recommended
that the ergoselect ergometer be
powered from an uninterruptible
power supply or a battery.
40 % UT
(60 % dip in UT) for 5 cycles
40 % UT
70 % UT
(30 % dip in UT) for 25 cycles
70 % UT
< 5 % UT
(> 95 % dip in UT) for 5 seconds
< 5 % UT
Power frequency
(50/60 Hz) magnetic
field to EN 61000-4-8
3 A/m
passed
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
The ergoselect ergometer has no
components susceptible to magnetic
fields.
Note:
UT is the a.c. mains voltage prior to application of the test level.