background image

- 12 -

Caution

•  Equipment Damage  •

Avoid strong vibrations of the ergoselect 50 during transport.

m

oUntinG

 

the

 C

ontrol

 t

erminal

The control terminal can be installed with the display 
either facing the patient or the operator.

It is recommended to install the terminal with the display 
and control keys towards the operator and the speed 
display towards the patient.

t

ransport

For short distances, the ergoselect 50 can be lifted at the 
saddle and rolled away on its castors.

To cover greater distances, however, we recommend the 
following method:

•  Disconnect the power cord from the wall outlet.

•  Rotate the handlebar towards the front. 

Tighten the clamping lever.

•  Stand in front of the ergoselect 50, grasp the handle‑

bar and tilt the ergoselect 50 towards you until it is 
standing on the castors only and is balanced.

•  It is now possible to transport the ergoselect 50.

•  When you have reached the new location, lower the 

ergoselect 50 very carefully to avoid damage.

t

ransporting

 

the

 

ergoselect

 50

M

ounting

 

options

 

for

 

control

 

terMinal

Summary of Contents for ergoselect 50

Page 1: ...Manual ergoselect 50 Bicycle Ergometer Operator s Manual 201000147000 Version 03 2010 English...

Page 2: ...d in this manual All trademarks appearing in this document are trademarks of their respective owners Their protection is acknowledged No part of this manual may be reprinted translated or reproduced w...

Page 3: ...ecting the Power Cord 13 Connecting the ECG Cable 15 Preparing the Patient 16 Adjusting the Saddle and the Handlebar 16 Operation Control terminal P 17 Turning the System On 17 Operating Modes with Co...

Page 4: ...4...

Page 5: ...5 Declaration of Conformity...

Page 6: ...6...

Page 7: ...st read and understand the content of this manual before using or work on it Paragraphs with special symbols are of particular importance If unauthorized individuals open the control terminal damaging...

Page 8: ...evice to other pieces of equipment Equipment Malfunction Cellular telephones may not be used in the immediate vicinity of the ergometer because they might interfere with the proper functioning of the...

Page 9: ...l If the ergometer is used for other purposes the manufacturer cannot be held liable for personal injuries or property damage resulting from the unintended use of the equipment Biocompatibility The pa...

Page 10: ...ybody connecting additional equipment to medical electrical equipment config ures a medical system and is therefore responsi ble that the system complies with the requirements for medical electrical s...

Page 11: ...t of the handlebar angle 3 Castors 4 Speed display RPM for patient 5 Adjustment of the saddle height 6 Power switch green button 7 Levelling devices to adjust the ergometer to uneven floors 8 Sockets...

Page 12: ...an be lifted at the saddle and rolled away on its castors To cover greater distances however we recommend the following method Disconnect the power cord from the wall outlet Rotate the handlebar towar...

Page 13: ...longer wobbles In case of delicate flooring it is recommended to place a mat under the ergometer to protect the flooring from damage by the feet The ergometer can be transported by means of the 2 fron...

Page 14: ...ide the power cord under the strain relief and screw the strain relief to the frame the plastic pin must lock into place in the boring provided Put the ergometer back on its feet and set the handle ba...

Page 15: ...face connector remote control through PC or ECG Connecting the ECG Cable ergoselect ergometers can be connected to electrocardio graphs and PC based ECG systems of most manufacturers Different connect...

Page 16: ...oximately 10 Set the handlebar to a position where the patient sitting upright on the saddle can reach it easily To do so open clamping lever 1 and set the handlebar to a suitable angle Adjusting sadd...

Page 17: ...option is selected the initial screen of the selected operating mode e g Ergometry will be displayed instead of the main menu With the key you can display the main menu If an error message e g E 01 a...

Page 18: ...load changes via the keypad SETTINGS Used to configure the ergometer Speed readout At the top of the control terminal there is a speed readout for the patient as well as three LEDs that inform the pat...

Page 19: ...he exercise test will start and the corresponding values will be displayed The exercise test can only be terminated with the corre sponding command from the controlling ECG unit Note All functions are...

Page 20: ...tion with SELECT The exercise test is started with the Start key a blood pressure measurement at rest may precede the test see Settings When the basic load appears on the display after approx 15 secon...

Page 21: ...e selected manually It is recommended that the patient continues to pedal in the recovery phase The END key in the middle will terminate the test End Note The saddle height ergoselect 200 can be chang...

Page 22: ...easurements The exercise test is started with the Start key afterwards the load can be set and changed with the 5 W and 5 W keys in increments of 1 W up to 25 W as config ured Blood pressure measureme...

Page 23: ...the ergometer will display this menu Use the softkeys on the right and left to position the bar cursor on your preferred default mode and save the selection with SELECT Protocols Protocols 6 15 are us...

Page 24: ...is adjustable in the range from 0 to 100 Load Change Here you determine the increments for each load change Depending on your choice each key press will change the load by 1 5 10 und 20 Watts Selecti...

Page 25: ...in with you select DATE or TIME and confirm the selection Then the value displayed in reverse video can be edited with the keys and saved with SELECT The time is adjusted in the same way You exit the...

Page 26: ...be initiated with digital commands Digital default The communication with the ergo meter is entirely controlled with digital commands Analog IN OUT The entire communication load control and BP mea su...

Page 27: ...e RPM limit values RPM Min 0 70 54 Max 50 130 64 Select Note The limits selected in this menu only apply to the load range between 6 and 150 watts At higher loads the RPM limits automatically adapt to...

Page 28: ...28...

Page 29: ...d with soap water The cloth should only be moist and not dripping wet If the cleaning agents and disinfectants used are caustic or contain alcohol they may damage and or discolor the upholstery Only t...

Page 30: ...o years according to the rules of the art by a Service Engineer authorized by ergoline Similarly the automatic sphygmomanometer in the control terminal must be checked and calibrated by an authorized...

Page 31: ...ad Increments user programmable Internal Protocols Control Terminal P 5 fixed incremental exercise test protocols e g WHO 10 user programmable protocols manual load control Permitted Patient Weight 16...

Page 32: ...on temperature 10 to 40 C 50 to 104 F rel humidity 30 to 75 no condensation atmospheric pressure 700 to 1060 hPa transport and storage temperature 40 to 70 C 40 to 158 F rel humidity 10 to 90 no conde...

Page 33: ...black speed independent range to DIN VDE 0750 0238 black grey speed independent range of the ergoselect ergometer Family of characteristics of load duration due to IEC 60601 1 Under permanent load co...

Page 34: ...34...

Page 35: ...operation in the configuration in which it is being used Guidance and Manufacturer s Declaration Electromagnetic Emissions The ergoselect ergometer is intended for use in the electromagnetic environm...

Page 36: ...differential mode 2 kV common mode 1 kV 2 kV Mains power should be that of a typical commer cial or hospital environment Voltage dips short interrup tions and voltage variations on power supply input...

Page 37: ...from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment mark...

Page 38: ...t Power of Transmitter W Separation Distance According to Frequency of Transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0...

Page 39: ......

Page 40: ...ergoline GmbH Lindenstra e 5 72475 Bitz Germany Tel 49 0 7431 98 94 0 Fax 49 0 7431 98 94 128 e mail info ergoline com http www ergoline com...

Reviews: