On the ergoselect 50, you adjust the saddle height manu‑
ally with a clamping lever.
With the pedal in the bottom position, the angle between
the axis formed by the upper body and the thigh should be
approximately 10°.
Set the handlebar to a position where the patient sitting
upright on the saddle can reach it easily.
To do so, open clamping lever 1 and set the handlebar to a
suitable angle.
a
djusting
saddle
and
handlebar
1
Adjustment of the handlebar angle
2
Adjustment of the saddle height
Note
• Tighten the clamping levers only as far as necessary, NOT
with maximum force.
• Lubricate the clamping lever threads quarterly at mini-
mum, using a suitable lubricant (e.g. OKS470).