Maintenance & Safety

 Cleaning and Maintenance

The following procedures are not guaranteed to control infection. The hospital infection control administrator or epidemiologist should be consulted regarding cleaning procedures 
and processes.

 To avoid risk of electric shock, do not expose electrical components to water, cleaning solutions or other potentially corrosive liquids or substances.

 Do not immerse Cart or Cart components in liquid or allow liquids to fl ow into the Cart. Wipe all cleaners off  surface immediately using a damp cloth. Thoroughly dry surface 

after cleaning.

 Do not use fl ammable cleaners on Cart surfaces due to close proximity of electrical power and equipment.

All paints and plastic Cart components will withstand cleaning by most commonly used, diluted, non-abrasive solutions such as quaternary ammonia compounds, ammonia 
enzyme cleaners, bleach or alcohol solutions.
• Pen and permanent and dry erase markers can be removed with 91% isopropyl alcohol and a soft cloth.
• Iodine stains can be removed with commonly used cleaners and a soft cloth.
• Never use steel wool or other abrasive materials that will damage the surface fi nish.
• Do not use strong solvents such as trichloroethylene and acetone. These solvents will damage the surface fi nish.
It is recommended that any cleaning solution be tested on a small, inconspicuous area to ensure surface is not harmed.

 Adjustment, Service, Replacement - DO NOT attempt to adjust, service or replace any part of the StyleView Cart unless directed to do so through Ergotron-approved 

documentation (i.e. installation instructions). Only Ergotron, Inc. or an Ergotron-certifi ed entity may adjust, service or replace StyleView Cart components. If any component on the 
Cart is missing or damaged, the Cart must not be used, contact Ergotron Customer Care immediately to request a replacement part.

 Cables - Keep cables neatly organized on the Cart (a variety of solutions are provided with your cart for this purpose). Excess cables should be routed away from moving 

components with cable clips. Review Cable Routing section of this guide, or contact Ergotron Customer Care for more information.

 Casters - Check casters periodically to make sure they are clean and free of debris that would prevent smooth travel. Avoid moving Cart across uneven, dirty or damaged 

surfaces. 

 Customer Equipment- Make sure equipment is balanced and mounted securely to Cart. Do not reposition Cart components on riser or tower unless instructed to do so in the 

installation instructions. Moving Cart components too high or too low on the Riser may create an unstable condition, leading to equipment damage or even personal injury. Contact 
Ergotron Customer Care for information about moving Cart components.

Earth Bond Test: To ensure safety grounding between the 
power cord ground connection and any accessible metal 
parts on the cart, although not mandatory, the following test 
is recommended to be performed every two years using a 
calibrated medical device safety analyzer. The procedure is as 
follows:

1. Disconnect the cart from mains power (unplug cart from the 

wall).

2. Remove power from internal power system outlets by 

pressing the power button on the Power System Interface.

3. Set up Earth Bond Test per the medical device safety analyzer 

instructions.

4. Connect the Cart power cord to the medical device safety 

analyzer.

5. Remove cover to access Lift Tension Adjustment point.
6. Attach second medical safety device analyzer probe to Lift 

Tension Adjustment point on Cart.

7. Perform Test (25 amps).
8. Ensure Earth Bond is less than or equal to 0.2-ohms.
9. Remove medical device safety analyzer connections to Cart, 

replace cover over Lift Tension Adjustment point, return the 
Cart to service. 

 
Insulation Resistance Test: To ensure mains power lines are 
adequately insulated from earth ground, the following test, 
although not mandatory, is recommended to be performed 
annually using a calibrated medical device safety analyzer. The 
procedure is as follows:

1. Disconnect the cart from mains power (unplug cart from the 

wall).

2. Remove power from internal power system outlets by 

pressing the power button on the Power System Interface.

3. Set up the Insulation Resistance Test per the medical device 

safety analyzer instructions.

4. Connect the Cart power cord to the medical device safety 

analyzer.

5. Perform test (500 V).
6. Ensure Insulation Resistance is greater than or equal to 1 

Mohms.

7. Remove medical device safety analyzer connections to Cart 

and return Cart to service.

Recommended Periodic Inspection and Maintenance

Component

Action

How often

By whom

ALD Batteries 
(SV31 only)

Replace 4 ea AA NiMH batteries

1

Replace when 
keypad #4 fl ashing

Any user

Ready On Batteries 
(SV31 only)

Replace 2 ea AA Lithium batteries

1

Replace when light 
on Ready On fl ashing

Any user

UI, Ethernet, USB 
cables

Inspect for wear, pinching, bad 
connectors

Monthly

Any user

Fan on side of CPU 
compartment

Inspect for dust at intake, vacuum 
as required using a vacuum cleaner 
that DOES NOT generate ESD 
(Electrostatic Discharge)

Monthly

Any user

Casters

Inspect for wear and debris

Monthly

Any user

Maximum Load

Inspect to ensure that maximum 
recommended loads are not 
exceeded

Daily

Any user

Power System SLA 
batteries

Replace lead acid batteries

1,2

When instructed 
by StyleLink or 
when battery 
runtime is 80% of 
original runtime

IT Personnel

Power Module

Inspect for dust at intake, vacuum 
as required using a vacuum cleaner 
that DOES NOT generate ESD 
(Electrostatic Discharge)

Monthly

IT Personnel

Battery harness

Inspect for wear, cracking, pinching, 
or other damage

Monthly

IT Personnel

Coiled Cord

Inspect for wear, damage, or 
stretching. 

Weekly

Any user

1. Please dispose of all batteries in accordance with local law
2. Always replace with similar size battery (33 A-h to replace 33 A-h) and always replace in pairs

Summary of Contents for StyleView SV32

Page 1: ...sito De senaste versionerna av installationshandledningen och nedladdning av programvaran för StyleLink finns på 最新版ユーザインス トールガイ ドとスタイルリンクソフ トウェアは次のサイ トでダウンロードできます 请从下列地址获取最新版本的用户安装指南和 StyleLink 软件下载 사용자 설치 안내서 및 StyleLink 소프트웨어 다운로드는 다음 웹 사이트에서 제공하고 있습니다 A B 1 2 3 4 5 6 3x 1x 1x 1x 1x 1x 1x 2x 1x 2x 1x 4x M4 x 10mm 4x M4 x 10mm 1x 1x M4 x 8mm M4 x 5mm 14mm 9 16 6 1 8 m EN 888 24 092 G 02 rev H 05...

Page 2: ...mérique de la classe A est conforme à la norme NMB 003 du Canada FCC Compliance Statement The cart has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and ca...

Page 3: ...0 4 Drawers 9 compartments SV32 81226 Input 120VAC 60 Hz 5 1A Output 120VAC 60 Hz 400VA 300W The cart and power system are certified to UL 60601 and CAN CSA C22 2 601 1 M90 6 Drawers 6 compartments SV32 82211 SV32 83211 Input 230V 50 Hz 2 7A Output 230V 50 Hz 400VA 300W The cart and power system are compliant with EN 60601 1 1 Drawer 6 compartments SV32 82222 SV32 83222 Input 230V 50 Hz 2 7A Outpu...

Page 4: ...o Eliminate Independent LCD Lift section 3 lbs 1 4 kg Worksurface 13 lbs 5 9 kg CPU Compartment CAUTION If the combined LCD and CPU weight is greater than 23 lbs 10 4 kg then the CPU must be mounted to the rear of the cart using the Universal CPU Holder accessory ordered separately see inset below 4 5 lbs 2 0 kg without insert 2 5 lbs 1 1 kg with insert Total weight capacity per drawer Weight Capa...

Page 5: ... 1x 2x 1x EN 888 24 092 G 02 rev H 05 11 Set up Connect Keyboard and Mouse to USB Hub NOTE Bar Code Scanner should be connected directly to computer USB port DO NOT connect Bar Code Scanner to the USB Hub USB Type A USB Type A ...

Page 6: ...d Moving Parts can Crush and Cut Raise monitor to top of vertical adustment BEFORE removing Failure to heed this warning may result in serious personal injury or property damage 822 310 With Independent LCD Lift 6 16 lbs 2 7 7 3 kg Without Independent LCD Lift 20 lbs 9 kg See How To Eliminate Independent LCD Lift section ...

Page 7: ...ra Small Form Factor CPU s Center CPU in storage area USFF Ultra Small Form Factor Thin Client Route computer cables down opening in enclosure left or right side Route computer cables down opening in enclosure left or right side Place AC Power Adaptors in compartment Route power cable down along tower Power cables need to be 6 1 8 m long ...

Page 8: ...8 21 d c 1x 1x e f a 5 EN 888 24 092 G 02 rev H 05 11 Set up ...

Page 9: ...hes when charging power cord plugged into wall outlet Allow battery to continue charging until light stops flashing After light stops flashing it is OK to unplug the power cord from the wall You can use cart while charging Battery has less than 90 charge Battery has less than 70 charge Battery has less than 50 charge Battery has less than 30 charge Light is red and alarm beeps Plug in power cord a...

Page 10: ...6 If computer still does not turn on contact Ergotron Customer Care Keyboard Light Keyboard Light will automatically turn off after 15 minutes if not manually turned off USBType A connection on computer Ethernet Crossover connection on computer NOTE these cables are factory connected to the USB Hub WARNING Fan must always be running when computer is on Operating computer without fan may lead to ov...

Page 11: ... load you should repeat these adjustment steps to ensure safe and optimum operation Adjustments should move smoothly and easily through the full range of motion and stay where you set it If adjustments are difficult and do not stay in the desired position follow the instructions to loosen or tighten the tension to create a smooth easy adjustment motion Depending on your product and theadjustment i...

Page 12: ...Lift 6 16 lbs 2 7 7 3 kg Without Independent LCD Lift 20 lbs 9 kg WARNING IMPACT HAZARD MOVING PARTS CAN CRUSH AND CUT DO NOT remove Stop Screw without monitor attached Doing so will cause monitor pivot to shoot up rapidly and may cause personal injury To Remove Monitor 1 Remove stop screw while monitor is attached 2 Lift monitor to highest postion 3 Remove monitor Failure to heed this warning may...

Page 13: ...rawers will light 2 Simultaneously press 4 and 6 Numbers 4 and 6 will light green 3 Simultaneously press 4 and 6 again Numbers 4 and 6 will flash green 4 Simultaneously press 4 and 6 a third time All numbers will light green You have 10 seconds to enter a new User PIN 5 Input new 5 digit User PIN All numbers will flash green You can continue to enter 5 digit User PINs until finished After sitting ...

Page 14: ...rs 1 Amp Fuse Turn power system off by holding down the AC Outlet Power button for 1 3 seconds Power light will shut off Turn off all mounted equipment Recommended Littelfuse 312001P Disconnect Power System from power source Before reseting circuit breakers contact Ergotron to determine cause of trip ...

Page 15: ...batteries while cart is located in an oxygen rich or hazardous environment arcing may occur and cause combustion Replace first battery before removing second battery to reduce the risk of cables touching terminals and causing arcing Always replace both batteries Replace first battery before removing second battery Replace batteries with same Amp Hour rating batteries only Only Ergotron specified b...

Page 16: ... Black Caution Connect Red before connecting Black Caution Remove Black before removing Red Red Red Black BATTERY LEAD Recycle battery or contact Ergotron for proper battery disposal guidelines Red Caution Connect Red before connecting Black Black Follow Battery Charge Discharge Initial Power on steps ...

Page 17: ...g Disconnect Power System from power source Once battery has been charged to 100 turn power system off a Turn power system off by holding down the AC Outlet Power button for 1 3 seconds Power light will shut off b Turn both circuit breakers to OFF NOTICE Warranty on fully charged batteries left in an unused state for more than three 3 consecutive months is automatically void Warranty on fully disc...

Page 18: ...ber charge battery fully 100 every day Maintenance Safety Color Symbol Signal Word Level of Hazard Red DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious injury Orange WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injury Yellow CAUTION Indicates a potentially hazardous situation which i...

Page 19: ...ower cord ground connection and any accessible metal parts on the cart although not mandatory the following test is recommended to be performed every two years using a calibrated medical device safety analyzer The procedure is as follows 1 Disconnect the cart from mains power unplug cart from the wall 2 Remove power from internal power system outlets by pressing the power button on the Power Syste...

Page 20: ...erential mode 2 kV common mode Complies Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 seconds Complies Mains power quality should be that of a ...

Page 21: ... mm 15 5 394 mm 1 8 46 mm 10 3 261 mm 4 102 mm 1 65 41 8 mm 16 8 427 mm 4 4 9 4 112 239 mm 24 2 44 2 614 1123 mm 47 4 67 4 1204 1712 mm 30 6 50 6 778 1285 mm 5 3 134 mm 4 4 112 mm 8 8 224 mm EN 888 24 092 G 02 rev H 05 11 Features Specifications When figuring dimensions include mounted accessories protruding cables and port replicators or docking stations CPU Compartment ...

Reviews: