7 | Control/operation
Erme AG
48 / 66
Table models professional
"General settings" menu
The following settings can be made in the "General settings" menu.
Menu item
Step
Setting range
Description
Service program
Starts the service program.
Follow the instructions shown on the display.
P calibration
Starts the pressure calibration.
Follow the instructions shown on the display.
Language
DE, EN, FR, IT
Selects the language between German, English, French and Italian.
Pump off delay
1
Off
1 - 254 s
Duration
After the vacuum procedure is completed, the pump is switched
off after the set time expires.
Standby after
1
Off
3 - 180 minutes
Default setting: 5 min.
The machine is automatically switched into standby mode if no
entries are made within the set time.
GreenVac option: this function is deactivated.
Max. Pump On t
1
Off
3 - 180 minutes
Default setting: 30 minutes
Maximum duration of pump operation. After the set time expires,
the machine switches the pump off.
This function protects the motor against overheating.
Time to service
Remaining time until the next recommended service. After this
time expires, a service reminder is shown on the display after the
machine switches off.