8. Maintenance
To ensure that
MyLab
operates over time at its maximum efficiency, Esaote
recommends to perform maintenance procedures regularly.
Maintenance procedures should be performed both by the user itself and by
Esaote authorized service personnel. The maintenance operations and
schedule are provided in the table below.
Table 8-1: Maintenance Operations
Maintenance
Minimum frequency
Performed by
cleaning the probes
after use
user
checking the probes
every week
user
cleaning control panel and device
every week
user
cleaning touch screen
every week
user
cleaning probe and gel holders
every week
user
cleaning LCD case and screen
every week
user
checking device housings for any
damage
every month
user
checking control panel and keyboard for
defects
every month
user
checking equipment for loose or missing
hardware
every month
user
checking movements of all parts
composing the device
every month
user
checking LCD and touchscreen status
every month
user
checking trackball movement
every month
user
cleaning trackball
every month
user
checking connectors on cables for any
mechanical defects
every month
user
Summary of Contents for MyLabX75VET
Page 1: ...Rev 03 June 2021 MyLabX75VET GETTING STARTED 350080800 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi ...
Page 61: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED ...