A D D I T I O N A L I N F O R M A T I O N O N S A F E T Y
MyLab
- G E T T I N G S T A R T E D
2 - 3
GETTI
NG STARTED
Electromagnetic Compatibility
This device was designed for use in the electromagnetic environments
declared in the tables below, in compliance with standard IEC 60601-1-
2:2014 (4
th
edition). The operator must make sure that s/he uses it in keeping
with this standard.
Ultrasound scanners require special precautions regarding EMC and must be
installed and put into service according to the provided information.
Ultrasound scanners are designed to generate and receive radiofrequency (RF)
energy and are, therefore, susceptible to other RF sources. As an example, other
medical devices, information technology products or TV/radio transmitters may
cause interference with the ultrasound system.
NOTE
Sensitivity to interference is more noticeable in Doppler modes.
In the presence of RF interference, the physician must evaluate the image
degradation and its diagnostic impact.
CAUTION
Use of accessories, probes and cables other than those authorized by
Esaote may result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation.
CAUTION
Portable RF communication equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of
MyLab
, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result.
WARNING
Portable and mobile RF communication equipment may cause
interference with the ultrasound system. Do not use these devices in
proximity of ultrasound equipment.
If an ultrasound system causes interferences (which can be identified by turning the
system off and on) with other devices, the user could try to solve the problem by:
relocating the system,
increasing the distance from other devices,
Summary of Contents for MyLabX75VET
Page 1: ...Rev 03 June 2021 MyLabX75VET GETTING STARTED 350080800 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi ...
Page 61: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED ...