Error Messages
Whenever an internal fault occurs, the device automatically freezes and an
error message is displayed on the screen. Switch
MyLab
off and then turn it on
again to see whether the error message persists.
AO
Save anyway the log file (refer to the “Archive” section of the Advanced
Operation manual for further information) and contact the Esaote Service
department.
Errors in Battery Management
The battery icon is shown crossed out whenever an error in the battery
management occurs.
The number in the warning message indicates the type of error.
Error #1
This error indicates a fault on the Power Supply: in this case, information on
batteries may not be correct.
MyLab
displays the following message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #2
This error indicates a failed access to the battery pack: in this case,
information on batteries may not be correct.
MyLab
displays the following
message:
If this situation occurs, shut down
MyLab
, by keeping
O N /O FF
pressed, and
contact Esaote Service.
Error #3
This error indicates that one battery couldn’t be charged.
MyLab
displays the
following message:
If this situation occurs, close the exam as soon as possible by pressing
EN D
EXAM
and switch
MyLab
off by pressing
O N /O FF
and then the main switch
Error #1: wrong communication with
power supply. The automatic shut down
is disabled.
Error #2: wrong communication with
battery logics. The automatic shut down
is disabled.
Error #3: problem with battery
charging
.
Summary of Contents for MyLabX75VET
Page 1: ...Rev 03 June 2021 MyLabX75VET GETTING STARTED 350080800 ...
Page 2: ......
Page 5: ...MyLab G E T T I N G S T A R T E D v GETTING STARTED RED Declaration of Conformity ...
Page 6: ...MyLab G E T T I N G S T A R T E D vi ...
Page 61: ...S Y S T E M O V E R V I E W MyLab G E T T I N G S T A R T E D 3 17 GETTING STARTED ...