14
Module IP Address
Command:
MOD:IP <nnn.nnn.nnn.nnn>
Description:
The device default IP address and subnet mask is 192.168.0.100, 255.255.255.0. The default
address and subnet mask are assigned to the device by ETS-Lindgren and do not change even
if your computer reboots. The IP address can be changed using the MOD:IP command. The new
address will not change even if your computer reboots. The port number is 1206.
Query:
MOD:IP?
Returns:
nnn.nnn.nnn.nnn
Example:
MOD:IP 192.168.0.55
Module Name
Command:
MOD:NAME <Module Name>
Description:
The <Module Name> parameter in the *IDN? query response is a place holder to identify a
specific device in a network. If you have more than one device you might want to identify them
with different module names. For instance, “EMC LAB1” and “EMC CHAMBER”.
Query:
MOD:NAME?
Example:
MOD:NAME EMC LAB1
Module Subnet Mask
Command:
MOD:NETMASK <nnn.nnn.nnn.nnn>
Description:
The device default IP address and subnet mask is 192.168.0.100, 255.255.255.0. This address
and mask are assigned to the device by ETS-Lindgren and does not change even if your
computer reboots. The subnet mask can be changed using the MOD:NETMASK command. The
new subnet mask will not change even if your computer reboots.
Query:
MOD:NETMASK?
Returns:
nnn.nnn.nnn.nnn
Example:
MOD:NETMASK 255.255.0.0