iMiniPlus PDF Logger User Guide v2.0 

 

 

- 17 - 

9.3.1  Specification and Alarm 

Raise alarm on below lower spec readings

 – tick this box if you want to activate the low alarm and select the value 

in the box immediately to the right. Note if the lower spec value is 0

C, then the alarm is triggered at -0.1

C, not 

0.0

C as this value is considered within specification. 

Raise alarm on above upper spec readings

 – tick this box if you want to activate the high alarm and select the 

value in the box immediately to the right. Note if the upper spec value is 50

C, then the alarm is triggered at 50.1

C, 

not 50.0

C as this value is considered within specification. 

Increment

 - the grouping size used for listing out-of-specification readings. 

Alarm Type 

This feature allows the user to specify the conditions that will trigger an alarm. 

Standard Alarm 

Number of consecutive out of specification readings

 – the alarm is triggered if the readings are consecutively over 

or under the specified threshold. For instance, if the upper spec is 50

C and the value for the number of consecutive 

out-of-spec readings is 3 and the logging interval is 10minutes, the readings will need to be in excess of 50°C for 30 
minutes before the alarm is triggered. If the readings fluctuate outside upper and lower threshold temperatures, 
the same rule applies (providing the appropriate “raise alarm” box is ticked). Once the reading(s) return inside alarm 
limits, the consecutive alarm is reset. 

Total number of out of specification readings (cumulative) –

 the alarm is triggered if the total number of out of 

spec readings equals or exceeds the entered value. In this case the alarm is NOT reset if readings return to within 
the specified parameters. 

Do not set the Total number of out-of-spec readings value to be less than the Consecutive out-of-spec readings as this 
will mean that the Consecutive out-of-spec readings alarm will never be triggered. 

Degree-Minutes Alarm (Integrator Alarm) 

The  Degree-Minutes  Alarm  is  a  special  feature  that 
“integrates”  temperature  and  time  parameters  into  one 
value. 

 

Total number of degree-minutes below specs exceeds

 - 

the  alarm  is  triggered  when  the  total  number  of 
temperature  degrees’  times  the  minute’s  duration  below 
the low alarm spec is greater than the value entered. 

Total number of degree-minutes above specs exceeds

 - 

the  alarm  is  triggered  when  the  total  number  of 
temperature  degrees’  times  the  minute’s  duration  above 
the upper alarm spec is greater than the value entered. 

Catastrophe Alarm 

This alarm is always additional to the Standard and 
Degree-Minute alarms and there is no delay. Do not set 
Catastrophe Alarm inside normal alarm limits because other alarms will never be triggered. 

Instantly raise alarm on readings below

 - tick this box if you want to immediately trigger the low alarm and select 

the value in the box immediately to the right. Note if the lower spec value is 0 °C, then the alarm is triggered at -0.1 
°C, not 0.0 C as this value is considered within specification. 

Summary of Contents for iMiniPlus PDF

Page 1: ...iMiniPlus PDF User Guide Version 2 0...

Page 2: ...6 9 3 1 Specification and Alarm 17 9 3 2 LCD Display 18 9 4 Checking the time clock 19 9 5 Setting the Start Stop and logging duration 19 9 5 1 Duration of trip at least 19 9 5 2 Interval between each...

Page 3: ...ning logger data to make a decision on accepting or rejecting the shipment 3 Logger profile The iMiniPlus PDF Logger is available in two models Both have the same temperature range of sensor D range 4...

Page 4: ...your distributor for a replacement When replacing the battery align one triangular mark on the battery cover against similar shaped mark on the logger base to avoid damaging plastic lock feature of th...

Page 5: ...r communication purposes with software LRF file is used by Console Pro to create secure LCF file for traceable data records iMiniPlus PDF Logger does not consume battery power when connected to USB an...

Page 6: ...r of the logger Hardware version this is actually firmware version Description string programmed to the logger by Console Pro Battery status last measured by the logger before it has been connected to...

Page 7: ...d on readings retrieved from logger memory and result may be different Lowest temperature reading for the current trip High Alarm status Low Alarm status Both High and Low Alarm can have the following...

Page 8: ...t to 20 C Note that the graph contains coloured dashed lines for primary alarm limits and solid coloured lines for catastrophe alarm limits Note that the numerical values of the alarm limits are not s...

Page 9: ...trip This part of the report can be disabled when programming You can change date display format to display month first by programming logger Date and Time is displayed along the X axis Temperature v...

Page 10: ...be read as Logger has spent 49 minutes and 40 seconds between 25 C and 25 9 C in this case the value of 1 degree per report line is configured by programmable parameter named Increment Note if Fahrenh...

Page 11: ...ways has two columns This part of the report can be disabled when programming You can change date display format to display month first by programming logger Marked readings are displayed with asteris...

Page 12: ...r The cable has two different ends The smaller one is type Mini B and the bigger one is type A Connect USB cable with Mini B smaller connector to the logger and A size to PC or laptop computer as illu...

Page 13: ...to use Console Pro Professional version 2 12 07 or later to program the logger Check this check box and click on Program and Configure button on the toolbar Program Wizard should start Then click Next...

Page 14: ...report that can be several pages long depending on the number of readings If you uncheck Create PDF file checkbox the logger will not generate PDF report at all You may need this option to prevent unw...

Page 15: ...year format by choosing one option from drop down box on this tab as shown below 9 2 Programming password The iMiniPlus PDF Logger supports a password option that when enabled results in users being...

Page 16: ...to add a password for the first time you can do this here Protect program settings Tick this option if you want to protect your programming settings for the current logger Protect retrieval of reading...

Page 17: ...eturn inside alarm limits the consecutive alarm is reset Total number of out of specification readings cumulative the alarm is triggered if the total number of out of spec readings equals or exceeds t...

Page 18: ...N 4 19 OFF 0 OFF 2 OFF 4 ALARM ON 5 21 OFF 1 OFF 3 OFF 5 ALARM ON 6 22 OFF 2 OFF 4 OFF 7 ALARM ON 7 24 OFF 3 OFF 5 ALARM ON 11 ALARM ON 8 22 OFF 4 ALARM ON 6 ALARM ON 13 ALARM ON 9 21 ALARM ON 5 ALARM...

Page 19: ...ent for instance you have the option of setting up the parameters yourself or let the Console Pro software calculate these automatically for you Thesoftware takes thetotal memory available 8048 readin...

Page 20: ...iMiniPlus PDF Logger User Guide v2 0 20...

Page 21: ...on the LCD You can also enable the STOP button on the logger This is disabled by default because in most applications it is better to take too many readings than too few Downloading the logger readin...

Page 22: ...utton If the upload has failed either select the Cancel button to close the software and start again or select the Restart button to start the sequence again If the iMiniPlus PDF Logger still fails to...

Page 23: ...ve the data from an iMiniPlus PDF Logger both of which will achieve the same result 1 Click the Download Information icon on the toolbar 2 Click the Download Information item located under the Action...

Page 24: ...application Otherwise leave the Logger option at its default setting Do not upload any new start conditions to the logger s Click on the Finish button Readings and Summary is displayed Download Wizard...

Page 25: ...d to communicate with the logger In other words you can email LRF files and Console Pro will create an LCF file from the LRF file Then you will be able to add digital signatures to the LCF file view d...

Page 26: ...n menu The following dialog should be displayed Please ensure that file name is unique By default Console Pro assigns the logger s serial number and trip number to the name of the file Navigate to the...

Page 27: ...and date Real time clock accurate to 1minute per month Sampling frequency 5 seconds 18hours Power source User replaceable 3V lithium coin cell Typical battery life 1 2 years operating life from a user...

Page 28: ...our website www cryopak com http www cryopak com en verification products Mailing Address PO Box 309 Buchanan VA 24066 Office 120 Parkway Drive Buchanan VA 24066 Phone 1 540 254 1433 1 732 346 9200 Ex...

Reviews: