Issue Date: 23 May 2011
Version 1.1
Escort iMiniPlus PDF Logger
Quick
User Guide v1.1
Escort Verification Technologies, Inc.
dba
ESCORT Data Loggers
PO Box 309, 120 Parkway Drive
Buchanan, VA 24066
Ph: 540-254-1433 Fax: 540-254-2433
- 11 -
Email:
techsupport@escortdataloggers.com
Page2 of PDF report
Out of specifications
Note: if Fahrenheit temperature unit scale is selected, the accuracy of “Out of specification” report
summary increment limits are ±0.1°F due to the complexity of conversion between Fahrenheit and Celsius.
The same report produced by Escort Console may differ from PDF report produced by the logger in term
of temperature by ±0.1°F. Duration of out of specification and the list of readings will read the same.
Page3 of PDF report
List of readings
Out of specification is a report of how long the logger spent outside of alarm limits
This part of the report can be disabled when programming
Each line in this report can be read as:” Logger has spent 49 minutes and 40 seconds between 25
°C and 25.9 °C”
In this case the value of 1 degree per report line is configured by programmable parameter
named “Increment”.
List of readings
•
This part of the
report always has two
columns.
•
This part of the
report can be disabled when
programming
•
You can change date
display format to display
month first by programming
logger
•
Marked readings are
displayed with asterisk sign.
•
If logger has 8048
readings, the list of readings
will be 73 pages long.