The table below demonstrates arbitrary data from an iMiniPlus
PDF Logger and how different alarm sources will be triggered.
This is complicated scenario for alarm settings based on:
Upper spec alarm: +20
°
C
Number of Consecutive out of spec readings: 4
Total number of out of spec readings: 5
Total number of degree-minutes above specs exceeds: 8
Instantly raise alarm on readings above: +23
°
C
10.3.2 LCD Display
The iMiniPlus PDF Logger LCD can be
configured to display or not display the
temperature information.
Enable Display
Tick this box if you want to enable the
display to show the temperature readings
and statistics.
If the box is un-ticked, all other features
will be grayed out. The iMiniPlus PDF
Logger will show “d-off” instead of a
temperature value. Additionally you will not
be able to scroll through the highest lowest,
average and out of spec readings. The word
“LATEST” will flash and if any alarm had
been triggered, “HIGH ALARM” or “LOW
ALARM” will be displayed.
The bookmarking function can still be used
when the display is disabled.
When the iMiniPlus PDF is stopped using the Stop button or it stops because it
has reached the end of a logging trip, the words “d-off” will disappear and the
word “STOPPED” will appear.
Display Units of measurement
You can select the LCD to display the temperature in Celsius or Fahrenheit.
Display change rate
You can select a frequency or 1, 2, 3 or 4 seconds that the LCD scrolls between
the Highest, Lowest, Average and out-of-spec information. Note that the
“LATEST”
reading will only change at a frequency equal to the logging interval.
So if the iMiniPlus PDF is programmed to log every 1 minute that will be the rate
at which the
“LATEST”
reading changes.
10.4 Checking the time clock
There maybe times when the time clock on the iMiniPlus PDF Logger may alter.
This is the case when the battery is removed and not replaced within 30seconds.