Issue Date: 23 May 2011 

Version 1.1

 
 

Escort iMiniPlus PDF Logger 

Quick

 User Guide v1.1 

 

 

Escort Verification Technologies, Inc.

 dba 

ESCORT Data Loggers 

 

PO Box 309, 120 Parkway Drive 
Buchanan, VA 24066 
Ph: 540-254-1433 Fax: 540-254-2433   

         

- 20 -

 

          Email: 

techsupport@escortdataloggers.com

 

 

   

11

 

Viewing logger data with PDF Reader Software 

 
You can download data from the logger by connecting the logger to a USB port. 
Once the logger hardware is installed and logger can be recognized by Operating 
System as Mass Storage Device you can open RECORD.PDF file. You can validate 
logger data by simply looking at PDF report. 
 

Please note that PDF Reader software is not able to save opened PDF file to 
other location if logger has been disconnected from PC after PDF report was 
opened with PDF Reader. 
 
Please note if you can not save PDF Report opened in PDF Reader software using 
“Save As” menu item you can always copy file using Windows Explorer and paste 
it to the desired location. 

 

12

 

Downloading logger readings via Escort Console Pro 

 
You can download data from the logger by connecting the logger to Console.  

Make sure the USB cable is connected to your PC and to the 

iMiniPlus PDF Logger. 

There are two ways to retrieve the data from an iMiniPlus PDF Logger, both of 
which will achieve the same result.  

1.

 

Click the Download Information icon on the toolbar  

 

 
2.

 

Click the "Download Information..." item located under the Action menu. 

 

 

The process of retrieving data stored within an iMiniPlus PDF Logger does not 
erase it from the memory. If the iMiniPlus PDF is currently recording while you 
retrieve the information, it will continue to record without interruption. 
The Download Information feature initiates the Download Wizard. 
 

Summary of Contents for MU-IN-D-16-L

Page 1: ...ification Technologies Inc dba ESCORT Data Loggers PO Box 309 120 Parkway Drive Buchanan VA 24066 Ph 540 254 1433 Fax 540 254 2433 Email techsupport escortdataloggers com Escort iMiniPlus PDF Data Log...

Page 2: ...report contents 9 10 Programming logger with Escort Console 12 10 1 Customizing PDF report and changing details 12 10 2 Programming password 13 10 3 Configure Sensors and Alarm Settings 14 10 3 1 Spec...

Page 3: ...1 0 F 40 F to 14 F 0 6 F 14 F to 158 F 0 6 F 14 F to 158 F Sensor response time INT T90 of 20 min in moving air INT T90 of 20 min in moving air EXT T90 of 5 min in moving air EXT T90 of 5 min in movi...

Page 4: ...Drive Buchanan VA 24066 Ph 540 254 1433 Fax 540 254 2433 4 Email techsupport escortdataloggers com 2 Getting Started What you need A B C D A Escort Console Pro Software Software Disk Software is also...

Page 5: ...Guide v1 1 Escort Verification Technologies Inc dba ESCORT Data Loggers PO Box 309 120 Parkway Drive Buchanan VA 24066 Ph 540 254 1433 Fax 540 254 2433 5 Email techsupport escortdataloggers com 3 Fron...

Page 6: ...ly using default USB drivers iMiniPlus PDF does not require any proprietary software installation to download data from logger after logging trip is finished iMiniPlus PDF normally generates 3 files o...

Page 7: ...Escort Verification Technologies Inc dba ESCORT Data Loggers PO Box 309 120 Parkway Drive Buchanan VA 24066 Ph 540 254 1433 Fax 540 254 2433 7 Email techsupport escortdataloggers com 2 Through the esc...

Page 8: ...e traceable and auditable data record The PDF report can be changed by the user This is contrary to the FDA 21 CFR Part 11 requirements to data security In order to produce secure and traceable record...

Page 9: ...ed ESCORT MPP drive 9 PDF report contents Open double click RECORD PDF You can disable PDF report generation by programming logger with Escort Console software You can also enable or disable certain p...

Page 10: ...tics part includes Highest temperature reading for the current trip Average temperature reading for the current trip Please note if the logger is configured for continuous logging logger memory is get...

Page 11: ...temperature by 0 1 F Duration of out of specification and the list of readings will read the same Page3 of PDF report List of readings Out of specification is a report of how long the logger spent ou...

Page 12: ...nfigure button on the toolbar Program Wizard should start Then click Next Logger should be recognized and displayed 10 1 Customizing PDF report and changing details Click on Next button You can see ed...

Page 13: ...rther access to the iMiniPlus PDF Logger When the password option is enabled a window similar to the following picture will be displayed allowing the user to enter the correct password The Program and...

Page 14: ...e alarm is triggered at 50 1 C not 50 0 C as this value is considered within specification Increment the grouping size used for listing out of specification readings Alarm Type This feature allows the...

Page 15: ...erature and time parameters into one value Total number of degree minutes below specs exceeds the alarm is triggered when the total number of temperature degrees times the minute s duration below the...

Page 16: ...ogger will show d off instead of a temperature value Additionally you will not be able to scroll through the highest lowest average and out of spec readings The word LATEST will flash and if any alarm...

Page 17: ...ge to the date and time on this computer option 10 5 Setting the Start Stop and logging duration 10 5 1 Duration of trip at least If you are using the iMiniPlus PDF Logger to record the temperature of...

Page 18: ...way to start an iMiniPlus PDF Logger is by using the START button With other Escort products it is possible to have a timed or date start as well as a switch start You can also program a delay into t...

Page 19: ...utinely download data from the iMiniPlus PDF Logger otherwise data may be lost Careful selection of the logging interval is required in this case It is possible to download at anytime while the iMiniP...

Page 20: ...was opened with PDF Reader Please note if you can not save PDF Report opened in PDF Reader software using Save As menu item you can always copy file using Windows Explorer and paste it to the desired...

Page 21: ...back one step in the step by step progress and Next takes you to the next step The Next button is preceded by the Download button The Help button will provide description and instruction about the st...

Page 22: ...254 1433 Fax 540 254 2433 22 Email techsupport escortdataloggers com Click on the Finish button 13 Does Console work with PDF file Escort Console neither opens logger PDF report nor creates one Escort...

Reviews: