The cabinet is made of stainless steel. Clean with a commercially
available cleaner for stainless kitchen appliances.
Modular Circuit Board
The VALE-1-14Q Unit model E888 is equipped with a modular
circuit board that is easily replaced if a problem occurs. If the ultrasonic
action in the tank stops, (and most of it is due to failures occurring in
the circuit board) call us and a replacement circuit board will be shipped
immediately. This way the need for shipping the unit back for repairs is
eliminated, and the disruption is minimal.
The circuit board is replaced as follows:
-Unplug unit from 115VAC outlet and drain liquid from tank.
-Remove screws, which hold back panel to cabinet and lay back panel
down.
-Loosen and remove the four wires connecting the circuit board to the
terminal connector.
-Circuit board can be removed from back panel by removing four holding
nuts.
-Fasten new circuit board to back panel with four nuts. Make sure
circuit board is properly spaced from back panel.
-Connect and tighten the four wires to terminal connector by matching
colored wires to corresponding color on the terminal block inside the
unit.
-Make sure wires from circuit board do not touch fan blades.
-Close back panel (first insert two screws in bottom of panel, then screw
tighten cabinet insides).
In case of any questions or problems contact the manufacturer.
Call 1-800-276-2466.
WARRANTY: One year
Two years on circuit boards
Lifetime on transducer bonds and tank weld seams.
Summary of Contents for E888