22
Is printer setting on?
Turn the printer setting on.
Is the error message displayed?
Follow the error message.
Are the printer
papers set correctly?
Set the printer paper correctly.
NO
YES
YES
NO
7-10
The measurement result is not printed out/printout is not normal.
The main causes when the measurement result is not printed out/printout is
not normal:
・
The printer setting is OFF.
・
Failure of the printer
・
Failure of the printer control board
・
Failure of the main control board
Replace the printer control board.
Is the printing normal?
Is the printer printing normal?
NO
YES
Replace the printer.
YES
NO
OK
NO
NO
Replace the main control board.
No print out/ printout is not normal.