56
12-8
Replacement of the roller, the measurement start switch board
ASSY, the joystick handle unit
1)
Press the claw of the stria part and remove the handle cover as pushing it up.
Stria
2)
Remove 2 harnesses that are connected to the measurement start switch board ASSY.
3)
Replace the roller and the measurement start switch board ASSY after unscrewing 2
screws that fix the board.
Roller
Measurement start
switch board ASSY
Fixing screw
Nut
4)
Replace the joystick handle unit after removing the measurement switch ASSY in
manner of 2) above and the nut.
RB-655-C02
At the time of reassembling, do not squeeze the harnesses between the surface of
the device and the nut and be careful about assembly direction of the resin
Left hand side of the examinee