58
10)
You can see the inside of the base if you lift the base cover 1 up. Unscrew 4 screws that
fix the mechanical unit ASSY and lift the mechanical unit up.
Mechanical unit ASSY
Fixing screws
11)
Remove the base cover 1 and set it as you can see the base unit.
12)
Replace PC I/F board, the board ASSY, and the power supply board after removing the
harnesses that are connected to the fixing screw.
Switching power supply PC I/F board ASSY
Terminal plate
Power supply board ASSY
Fixing screws
13)
At the time of replacement of the switching power supply, remove all of the harnesses
that are connected to the switching power supply after unscrewing the fixing screws of
the terminal plate and removing the terminal plate.
14)
The base cover will be removed when you unscrew 4 fixing screws.
15)
Replace it after unscrewing 2 fixing screws of the switching power supply.
Fixing screws
Fixing screws
The mechanical unit is very unstable so be careful when you remove it.
Fix the removed mechanical unit not to fall down.
Joyst
ic
k
si
de
Summary of Contents for AKR 750
Page 1: ...Auto kerato refractometer Maintenance Manual V1 09 2016 ...
Page 6: ...4 Block Diagram ...
Page 7: ......
Page 8: ...8 ...
Page 66: ...65 ...