45
11
Evaluation of Freeze Image
11-1
Normal condition
If the device is normal and it is aligned correctly, you see
the image screen as shown on the right.
11-2
In case that there is no measurement image
If you do not see the measurement image as shown on the
right after taking a measurement and switching to the
Freeze Image, the measurement cannot be taken. In this
case, there is a possibility that the measurement light is not
applied correctly.
First, check if the measurement light is applied or not
with following the procedure below.
Procedure: Return to the MAINTENANCE MODE screen and select ‘REF MEASUREMENT
LED’ in ‘5. Hardware Check’. If you can see the measurement light written above is
blinking on the screen or from the measurement window, it can be judged that the
measurement light is applied correctly. (Please check this with good alignment
condition.)In this case, it can be thought that the control board has some troubles.
Replace the control board. If you cannot see measurement light written above, it
can be though t that the measurement light drive circuit has some troubles. Check
with following the Troubleshooting 7-12.
11-3
In case that there is a measurement image
Even though you can see the measurement image after taking a measurement and
switching to the Freeze Image, there are some cases that you cannot take measurement.
Ex)
a) Unnecessary reflection of a cornea
b)
A measurement image is cut.
c)
A measurement image is blurred.
etc. can be thought. Specific examples are shown below.