background image

9

Molift Air 

Transport and Operating Conditions

Transport and Storage

For long time storage it is recommended that the 

emergency stop button is activated (pulled out). 

The lifter can be stored and transported under 

temperatures between -25 - 70 °C.

Operating

The lifter is designed for use at standard room 

temperatures (+5 to +40°C).

Air Pressure: 70 - 106 kPa

Relative Humidity: 15 - 93 %

Following storage or transport at other tempera-

tures leave the lifter in a room with a suitable 

temperature until it reaches a safe operating 

temperature.

Medical electrical equipment 

requires special precautions regard-

ing electromagnetic compatibility 

(EMC). Portable or mobile radio 

communication equipment may 

affect the medical electrical equip-

ment, and should be kept minimum 

25 cm (10 inches) from the lifters 

electronics.

Batteries

Molift Air has a 26.4 V NiMH battery pack. The 

battery has a life expectancy of approx. 500 

charge cycles.

Used batteries must be disposed 

of as special waste according to 

local rules and regulations. Do not 

dispose in general waste

x2

Batteries must be fully charged and 

depleted completely a couple of 

times before it achieves full capac-

ity. Also follow this procedure if 

the lifter has not been in use for a 

long period (3-4 weeks)

Battery and Service indicator

The lifter has two indicator lights.

Wrench symbol / Service light

Battery symbol / Battery light

The electrical system has a power save function

which will turn off the electrical system after ten

minutes without activity. All lights will turn off.

The system is activated when pushing one of the

operating buttons

Battery Light

The battery indicator light will illuminate and 

make a sound when battery level is low and the 

lifter requires charging. When this occurs the lifter 

will have sufficient power available for one full lift-

ing cycle with max load. When battery is critically 

low it is only possible to lower the lifting belt.

Service Light

The lifters electronics record the loads lifted and 

number of lifts. After a certain period of operation 

a signal is given to indicate that service is required.

Service light

Mode

No light

Power saving (Stand by)

Green

Ready for use

Yellow

Order service

Red

Perform service

Red + sound

Perform service

immediately

Service scope

Service involves replacing the lifting belt and 

inspection/replacement of worn parts. This must 

be carried out by authorized personnel.

Service is needed when the service light is red 

(calculated 5.000 lifts in weight class 2).

Electronics

Summary of Contents for Molift Air 205

Page 1: ...Molift Air EN User manual BM16101 Rev G 2015 12 07 ...

Page 2: ...In Rail Charging 10 Transfer 11 Using Slings 11 Lifting and lowering 11 Accessories 12 Suspensions 12 Slings 12 Scale 12 Maintenance 13 Cleaning and disinfeciton 13 Reconditioning 13 Recycling 13 Spare Parts 13 Periodic Inspection 13 Pericodic Inspection Scope 13 Troubleshooting 14 Important This User Manual contains important safety instructions and information regarding the use of the lifter and...

Page 3: ...and gait training situations together with the comprehen sive accessory program from Molift Molift AIR is a very light motor which in combina tion with the quick release coupling system gives an easy handling when mounting and servicing About Molift Air Hoist components 1 Trolley 2 Hoist 3 Side covers 4 Emergency stop Emergency lowering 5 Lifting belt 6 Quick release pin 7 Hand control 8 Suspensio...

Page 4: ...g Molift components and spare parts Declaration of conformity is not valid and Etac is not responsible for warranty if any modifications are made to the product Etac shall not be liable for faults or accidents that can occur when using components made by other manufacturers Only certified personnel are allowed to open hoist or accessories to perform service or repair Risk of injury from rotating p...

Page 5: ... 8kg 475 8lbs 26 4 V DC IP24 Duty cycle 10 Molift Air 300 Gross weight 310 8kg 685 2lbs 26 4 V DC IP24 Duty cycle 10 60601 1 60601 1 Warning labels and Symbols Symbols used on the product explained in more detail Class II double insulated CE marked Refer to user manual Indoor Use only 255 560 Max user weight Do not dispose in general waste Emergency stop Emergency lowering Service Light Battery Li...

Page 6: ...h 75kg 165 35 lbs load Propulsion motor speed 0 2m second 7 87 inches second Protection class Hand control IP24 Lift motor IP24 Maximum A weighted sound power level LWA 63 dB Operating forces button Buttons on handset 3 4 N Material Aluminum Plastic Expected Lifetime The hoist has an expected lifetime of 30 000 cycles or 10 years Dimensions L x W x H Length Width Height without trolley 36 x 19 x 1...

Page 7: ...ley No tools required The lifter has two connection points Connect one point first Push button all the way in and place lifter in connection point on trolley Release button and make sure it is clearly show ing green before connecting the next point Make sure both connection point buttons completely return and clearly show green after installation Emergency lowering cord and tube can be adjusted to...

Page 8: ...rd ISO EN 10535 Safe Working Load If maximum load SWL differs between hoist suspension and body support unit then the lowest maxi mum load shall always be used Molift lifters shall only be used to lift persons Never use the lifter to lift or move objects of any kind Inspection to be performed daily or before use Inspect lifter and rail system has no visible damage defects or deformations Make sure...

Page 9: ...ng Pull and hold to start lowering Hold until user is lowered and can be released from suspension Only for emergency use when hand control does not work Stop before full band is out or lifting band will be winded the wrong way and go upwards Manual Emergency lowering Allen key or bit for power drill can be found inside the lifter Remove cover to locate Push down and press cover outwards with a fla...

Page 10: ...full capac ity Also follow this procedure if the lifter has not been in use for a long period 3 4 weeks Battery and Service indicator The lifter has two indicator lights Wrench symbol Service light Battery symbol Battery light The electrical system has a power save function which will turn off the electrical system after ten minutes without activity All lights will turn off The system is activated...

Page 11: ... Trickle charge Red Green Error IRC In Rail Charging The lifter is charged through a built in charger in lift motor for continuous charging through the entire rail system length Molift Air IRC has a built in charger connected to the battery It must be installed with IRC Trolley in a rail system with a conductive tape The IRC trol ley is connected to the battery charger inside the Molift Air IRC Th...

Page 12: ... the user stand to the side of the person you are lifting Make sure that arms and legs do not obstruct the seat bed etc Keep eye contact with the user to help them feel safe The 4 point suspension must always be positioned across the user to prevent the user from uninten tionally sliding out of the sling The suspensions hook design prevents sling to be inadvertently detached when mounted correct 1...

Page 13: ...x with a single touch Read the manual that comes with the weight scale prior to use Slings Molift supplies a wide selection of slings for dif ferent types of transfers The Molift RgoSling sling series is developed to be combined with a 4 point suspension but the sling also works with a 2 point suspension See the combination list in the slings user manual for the correct sling and suspension combin...

Page 14: ...ccording to Periodic Inspection Report for Molift Air Periodic Inspection shall be performed at least once a year or more frequently if required by local requirements The inspection must be performed by service personnel authorized by Etac Contact Etac for training and authorization or recommen dation of an approved service partner When performing a periodic inspection the inspector shall fill out...

Page 15: ...roperly Open cover and plug in Hand control Hand control or plug or cord can be broken an should be replaced The lifter does not respond to Hand control up button Lifting strap angle sensor is activated Adjust lifting strap or move hoist to reduce lifting strap angle When using Electrical Emergency lowering the lifting band goes upwards Lifting band has gone completely out and has been winded the ...