Power Supply
The product is powered by theES5300.1-A Housing or the ES5300.1-B Housing
via the PCIe Backplane Connector.
Insulation Requirements for Lab Power Supplies to Circuits Connected to the HIL
System:
• The power supply to live circuitry must be safely isolated from the supply
voltage. For example, use a car battery or a suitable lab power supply.
• Only use lab power supplies with dual protection for the supply network
(with double/reinforced insulation (DI/RI)). This requirement is met by lab
power supplies that comply with IEC/EN 60950 or IEC/EN 61010.
• The lab power supply must be approved for use at a height of 2000 m and
in ambient temperatures of up to 40 °C.
De-energizing a Plug-in Board
Switch off the ES5300.1-A Housing or the ES5300.1-B Housing and external
power supplies, and unplug the power plug and other connectors attached to
the plug-in board. Wait at least three minutes before removing the plug-in
board.
WARNING!
Danger trough electromagnetic radiation!
The ES5372.1-B, fitted with the ES4455.2 or the ES4456.2 and loads
connected to it, can emit electromagnetic radiation during operation
which can cause interference with the operation or damage pace-
makers or implanted defibrillators.
The ES5372.1-B may be operated only in areas where persons with
pacemakers are prohibited from entering. At the entrances to these
areas, the P007 identification, "No access for persons with pacemak-
ers or implanted defibrillators" in accordance with ISO 7010:2011
"Registered Safety Signs", must be attached and clearly legible.
Failure to observe it can lead to health hazards and even death for
persons with pacemakers and implanted defibrillators.
WARNING!
Fire hazard!
Use only fuses that meet the specification in Tab. 2-1 on page 24!
Never bridge defective fuses!
Failure to observe the fuse specification can lead to excess currents,
short circuits and fires.